13
Analog Interface
Functions
– Auto-sensing mode
– 2 programmable trigger inputs
– 3 programmable signal outputs
– Voltage control input (0 ... 5 V)
– Current control input (0 ... 5 V)
– Voltage monitor output (0 ... 10 V)
– Current monitor output (0 ... 10 V)
– Master-slave parallel operation
– Master-slave series operation
– Auxiliary power output: 15 V / 60 mA
Computer Interfaces
•
IEC-625 / IEEE 488 interface (optional)
•
RS 232 interface
Transmission mode
asynchronous
Transmission speed
1200 to 115,200 baud, adjustable
•
USB port
USB port: 4-pin, type B
USB 1.1 compatible with USB 2.0
Connector pin assignments 1: VCC, 2: D-, 3: D+, 4: GND
Transmission speed
9600 to 115,200 baud, adjustable
Power supply
Line voltage
115/230 V ~ + 10 / – 15%; 47 to 63 Hz
Inrush current
Max. 50 A
S
Mains fuse
SYSKON P500/P800/P1500:
1 x M 15 A / 250 V (6.3 x 32 mm), UL
SYSKON P3000/4500: 3 x M15 A/250 V
Electrical Safety
Safety class
I
Measuring category
II for mains input
I for output and interfaces
Fouling factor
2
Earth leakage current < 2.5 mA
RMS
Electrical isolation
Test voltage
Output – mains
2.2 kV ~
Output – bus/ground 1.4 kV ~
Mains – bus/ground
2.2 kV –
Bus – ground
No electrical isolation
Applicated Standards
IEC 61010-1:2010
DIN EN 61010-1:2010
VDE 0411-1:2011
EN 61326
4.1.1
Electromagnetic Compatibility
SYSKON P500/P800/P1500
Generic standard
EN 61326-1: October 2006
Interference emission EN 55022: class B
Interference immunity EN 61000-4-2: feature A
EN 61000-4-3: feature B
EN 61000-4-4: feature A
EN 61000-4-5: feature A
EN 61000-4-6: feature A
EN 61000-4-8: feature A
EN 61000-4-11: feature A
SYSKON P3000/4500
Generic standard
EN 61326-1: October 2006
Interference emission EN 55022: class A *
Interference immunity EN 61000-4-2: feature B
EN 61000-4-3: feature A
EN 61000-4-4: feature B
EN 61000-4-5: feature B
EN 61000-4-6: feature A
EN 61000-4-8: feature A
EN 61000-4-11: feature B
*
Note:
Approved for the deployment in industrial environment. The
device may cause radio interferences in domestic areas.
4.1.2
Ambient Conditions
Temperature range Operation:
0 to 40° C
Storage:
–25 to +75° C
Atmospheric
humidity
Operation:
75% rel. humidity,
no condensation allowed
Storage:
65% rel. humidity
Cooling
With integrated fan
(temperature controlled)
Inlet vent:
Side panel
Outlet vent:
Rear panel
Operating noise
Noise pressure level at a distance of 30 cm
with fan set to low / high
Front
17 / 28 dBA
Rear
22 / 32 dBA
Left
17 / 28 dBA
Right
20 / 31 dBA