52
METRACELL BT PRO
Gossen Metrawatt GmbH
19a
A battery database has been created
18.
The battery database has been transferred to the battery tester
You use RFID technology for your system.
The DUTs in the battery database are linked to the corresponding RFID tags
1. Select
BATT DATAB.
with the
▼
key.
2. Press the
OK
key
The battery database is displayed. An acoustic signal indicates that the RFID reader is
ready.
3. Read the RFID tag (
An acoustic signal confirms reading and the DUT is displayed.
4. Press the
OK
key.
The battery is selected. The battery tester returns to the
MEASUREMENTS
menu.
START
->
BATT DATAB.
19b
A battery database has been created
18.
The battery database has been transferred to the battery tester
1. Select
BATT DATAB.
with the
▼
key.
2. Press the
OK
key
The battery database is displayed. An acoustic signal indicates that the RFID reader is
ready.
3. Select a battery using the
▼ ▲
keys.
4. Press the
OK
key.
The battery is selected and the battery tester returns to the
MEASUREMENTS
menu.
START
->
BATT DATAB.
19c
1. Select the
LOCATION:XXXX.
parameter with the
▼
key.
2. Press the
OK
key.
The entry mode is active.
3. Enter the parameter’s first digit
4. Acknowledge your entry with the
OK
key.
5. Repeat steps 1 through 4 for all other parameter digits for battery number, number of
blocks and upper and lower voltage limit values.
BATT DATAB.
->
LOCATION
:
0000
20
Return to
START
using the
▲
key.
->
START
BATT DATAB.
21
Press the
OK
key.
The battery tester is waiting for data.
WAITING FOR DMA 35
22
Select the following function sequence at the
DMA 35
:
Data Memory
Export Measurement Data
OK
.
A security prompt appears.
Export
Export all data?
23
Acknowledge the
Export all data
prompt with
Yes
.
24
The data are transmitted.
WAITING FOR DMA 35 (re-
ceiving data).
DMA 35: Export data
25
Data are transferred from the DMA 35 to the battery tester.
The measured value evaluation is displayed at the battery tester
Quantity:
[XXXX]
XX.XXXXV
MEAN:
+XX.XXXXV
DMA 35:Export finished
26
Select the following keys at the density meter
OK
-> Back -> Back - Back.
The initial screen appears.
27
Press the
OK
key repeatedly in order to view the evaluation (scroll), and to subsequently return
the display to
DMA35 (BT)
measurement.
QUANTITY:
[XXXX]
MEAN:
X.XXXkg/l
DMA 35:Export finished