44
METRACELL BT PRO
Gossen Metrawatt GmbH
9
Press the
OK
key in order to start the measurement.
BLOCK
- - - - - - -
XXXX/X
AN:XXXX
BN:XXXX
10
Contact the connector: using suitable pressure, press the multimeter test probes against the
connection points.
Example:
BLOCK
2513.15mV
0004/2
AN:0001
BN:0010
11
Wait until the measured value settles in.
12
Press the
OK
key in order to acknowledge the measured value.
The measured value is saved.
13
Remove the multimeter probes after hearing the acoustic signal.
14
Repeat the previously described measuring procedure for further connection points.
The measurement is ended automatically after all connection points have been completely
measured. The measured values are saved.
The measured value evaluation appears at the display
BLOCK
- - - - - - -
XXXX/X
AN:XXXX
BN:XXXX
15
Press the
OK
key repeatedly in order to view the evaluation (scroll), and to subsequently return
the display to the
CONNECTOR
measurement.
NUMBER
[XXXX]
AVER.:
XXXX.XXmV