51
Gossen Metrawatt GmbH
METRACELL BT PRO
10.12 DMA35 Measurement (BT)
No measurements can be performed with the battery tester itself by means of this measuring function.
Instead, measured density and temperature values acquired by and stored in the portable DMA 35 density meter (version 4) from Anton
Paar are transferred to the battery tester.
Note
Read the documentation included with the DMA 35 density meter from Anton Paar GmbH.
Data are transferred via Bluetooth
®
.
A CSV file is transferred from the DMA 35.
The CSV file uses different characters as field separators and decimal characters.
(Use the default values as settings for the DMA 35.)
Se-
quence
Description of the Measurement
Display
1
Press and hold the
FUNCTION
key for about 2 seconds in order to switch the battery tester on.
BTpro
<date>
<version>
<time>
2
Press any key to open the
main menu
.
If you’re already in another menu, press the
FUNCTION
key repeatedly until the
main menu
ap-
pears.
->
MEASUREMENT
DATA
3
Select
SETTINGS
using the
▼ ▲
keys.
DATA
->
SETTINGS
4
Press the
OK
key.
The
SETTINGS
menu is displayed.
SYSTEM
->
MEASUREMENTS
5
Select
BT SETTINGS
using the
▼ ▲
keys.
MEASUREMENTS
->
BT SETTING
6
Press the
OK
key.
LINK COMPUTER
->
LINK HEADSET
7
Select
LINK DMA35 V4
using the
▼ ▲
keys.
LINK HEADSET
->
LINK DMA35 V4
8
Press the
OK
key.
The battery tester can now be detected by the density meter.
PLEASE SEARCH WITH
DMA BTpro.
9
Select the following function sequence at the density meter:
Menu
Settings
Data Transmis-
sion
Configure Export Destination
OK
Start
.
A search for the export destination is started.
1: Start search.
2: Activate Bluetooth.
3: Search for devices.
4: Select device.
10
Press the Edit key in order to select
BTPRO
from the
Select Device
list.
Connection is established.
BTPRO
▼
11
Acknowledge your selection with
OK
.
4: Select device.
BTPRO
▼
12
Press
Next
.
5: Save your selection.
13
Acknowledge with
OK
.
The selection is saved and the
Data Transmission
menu appears.
Configure export destina-
tion
Configure printing destina-
tion
Configure printer
Configure data format
14
Select the
Back
Back
function sequence.
The
Menu
menu is displayed.
Data storage
Probe IDs
Methods
Settings
15
Go to the main menu at the battery tester.
->
MEASUREMENT
DATA
16
Press the
OK
key in order to open the
MEASUREMENTS
menu.
The
MEASUREMENTS
menu is displayed.
->
MULTIMETER
FLOAT
17
Select
DMA35 (BT)
using the
▼ ▲
keys.
DMA35 (IrDA)
->
DMA35 (BT)
18
Press the
OK
key.
->
START
BATT DATAB.
19
Before starting the measurement, the location, battery and limit value parameters must be set.
These can be:
–
Selected in the battery database via RFID
step 19a
–
Selected manually in the battery database
step 19b
–
Entered manually at the battery tester
step 19c
START
->
BATT DATAB.