INSTALLATION OF EQUIPMENT
Special providences or particular conditions for installation
- Check that the socket in which the device will be connected has a ground connection.
According to the ABNT standard, this is essential for the safe operation of the system.
- Install the unit in a place where it will not be damaged by the pressure, temperature,
humidity, direct sunlight, dust, salts.
- To guarantee an effective ventilation, the compressors should be installed or put with
airing grid away at least 20 cm of any obstacle that can obstruct the air passage, in a way
to facilitate the cleaning and maintenance operations.
- The unit should not be submitted to inclination, excessive vibrations, or blows (including
during transportation and handling).
- Install the equipment on a firm and leveled surface.
- This equipment was not planned for use in an environment where vapors, anesthetic
mixtures infl ammable with air, or oxygen and nitrous oxide can be detected.
- Check the voltage of the equipment at the moment of executing the electrical installation.
Electric installation
Breaker dimensioning
Wire gauge dimensioning
BioQualyAir 40L
BioQualyAir 65L
Wire diameter (Gauge):
For the distance up to 5m of breaker until the outlet of the
equipment, the wire should have a section 2,5mm², from 5 to 15m wire 4mm and from 15
to 50m wire 6mm (As by standard of ABNT NBR-0148, metric series PVC 70°C).
Voltage
Nominal Current
Breaker
127V~ ±10%
7,3 A
10 A
220V~ ±10%
5 A
10 A
Voltage
Nominal Current
Breaker
127V~ ±10%
14,6 A
20 A
220V~ ±10%
10 A
16 A
- It is compulsory the installation of a exclusive breaker in the distribution box for the
outlet where will be connected the compressor;
- Never uses adapters, extensions or transformers to link the compressor to the electric
net.
EARTHING OF THE ELECTRIC NET AND OF THE EQUIPMENT:
WARNING!
Inadequate connection of the earth wire can result in electric shock risk.