REPAIRS
03
Replacing the pressure manometer - output air slide valve -
pressure regulator
Replacing / repairing the solenoid valve(s) – overload
protector - capacitor(s)
- Remove the pressure regulator with filter (03) and provide
the substitution.
To make the maintenance procedures in the components: ”capacitor, solenoid
valve and overload protector” it will be necessary the removal of one of the motors.
IMPORTANT:
After the repair /substitution procedure
of the components, when mounting them again, it is of
extreme importance that it is guaranteed the complete
points sealing, using activator and lock thread glue.
Used tools:
- Fixed key 7/16”
- Fixed key 1”
- Fixed key 11/16”
- Fixed key 14mm
- Activator Loctite 7471
- Thread lock “Loctite 242”
- Loosen the hose with steel mesh of the motor output head (01) and input of the retention
valve (02);
- Loosen the air plastic hose unscrewing the squeeze nut (03).
- Loosen the electric command
box removing the four screws
(04);
- Loosen the motor loosening
the four screws that hold it.
(05).
01
03
04
05
02