Replacing / repairing the solenoid valve(s) - overload
protector - capacitor (s)
Capacitor
21
OBSERVATION
:
- Disconnect the capacitor terminals, remove from it’s support
and provide the substitution (21).
- Capacitors used in the compressors 220V/50Hz and 220V/
60Hz are of 25 μF or 35 μF.
- Capacitors used in the compressors of 127V/60Hz are of
85 μF.
After making any maintenance procedure in the components: ”solenoid valve
overload protector and capacitor” the electric connections should be accomplished
as the connection diagram (page.24).
Used tools :
- Fixed key 14mm
- Fixed key 10mm
- Fixed key 1/2”
- Fixed key 11/16”
- Phillips key
- Crimping plier
- Activator Loctite 7471
- Thread lock “Loctite 242”
REPAIRS