Note: If you are not able to solve an occurring problem,  

contact Biocomfort Diagnostics GmbH & Co. KG.

7 Technical data

Device type: 

Blood glucose meter 

Model: 

 gluco-comfort BGM105/ 

gluco-comfort BGM205

Serial no.: 

 Specified on the type 

plate (bottom of measur-

ing device)

Measuring  

method: 

 Glucose is oxidised by  

a glucose dehydrogenase, 

amperometric biosensor

Sample: 

fresh capillary blood

Sample volume: 

< 2 µl

Measuring range:   20 – 600 mg/dl 

(1.1 – 33.3 mmol/l)

Measuring period:  10 seconds

Coding: 

Code chip

Measured value  

memory: 

 Max. 122 blood glucose 

measurements per user, 

all values are stored with 

the time and date.

Average values: 

 7-, 14-, 21-, and 28-day 

average

User management: 

 1 to 8 users

Size: 

 70 x 90 x 18  

(W x L x H) [mm]

Weight: 

 circa 70 g  

(without battery)

Colour: 

 BGM105: Housing 

is grey aluminium, 

buttons and battery 

compartment cover are 

black 

BGM205: Housing is 

slate grey, buttons and 

battery compartment 

are light blue

Automatic switch-off  

function: 

 4 minutes when test 

strip is inserted; 

2 minutes if no func-

tion has been activated 

and no test strip is 

inserted.

Te

ch

ni

ca

l d

at

7

Error 

message

Causes

Measures

Remarks

23 error
(BGM 105)

The wireless connection 

could not be cut.

Switch off the device and 

restart.

Measurement cannot be carried 

out.

24 error
(BGM 105)

Communication error

Only occurs with the software, 

which is available as an acces-

sory. 

32 error

Function test of the system 

was not completed success-

fully. 

Switch off the device and 

restart.

If this error message appears 

repeatedly, have the device  

checked by the manufacturer.

error and 

blinking bat-

tery symbol 

Battery is almost completely 

used up.

Exchange the battery and 

restart the device.

When the battery symbol blinks, 

the blood glucose meter cannot 

be used.

Summary of Contents for BGM 105

Page 1: ...gluco comfort Blood glucose measuring systems BGM105 BGM 205 Operating manual Issued in 11 2006 REF 590107...

Page 2: ...the blood glucose measurement 28 3 4 Carrying out the blood glucose measurement 31 3 5 Use by multiple persons 35 4 Measured value memory 36 4 1 Storing the measured values 36 4 2 Calling the stored m...

Page 3: ...attery blinking dead battery 5 Number eld for Year Time User activated not activated Wireless connection activated not activated 6 Symbol for active wireless connection BGM 105 7 Code no indication wh...

Page 4: ...is operating manual contains important information for you It is mandatory that the safety measures mentioned in this operating manual be taken The Biocomfort blood glucose measure ment system may onl...

Page 5: ...ons for the test strips The test strips could be damaged by effects of temperature and light For this reason they should be used immediately after being removed from the test strip con tainer The test...

Page 6: ...tion regarding the control solution Please observe the separate user instructions for the control solution Always shake the control solution well before use The control solution can be damaged if expo...

Page 7: ...e Step 6 Afterwards check whether the battery has been correctlyinsertedbybrie ypressingtheOn Off button Inthiscase allelementsofthedisplayap pear for about two seconds as a function check Note Please...

Page 8: ...lue you desire Now you can go to the next parameter by pressing the SET button or you can exit the settings menu by pressing the On Off button Note If you keep the MEM button pressed you can change th...

Page 9: ...acti vate the minute display Step 10 Keep pressing the MEM button until the cor rect minute is displayed Step 11 If you are nished making settings you can exit the settings menu by pressing the On Of...

Page 10: ...1 Con guration of the model variant BGM205 In its delivered condition the measured value memory is only enabled for one user User 1 The con guration of the measured value memory can be found after th...

Page 11: ...ou wish you can set the status for user 2 by pressing the SET button Remark If for example you want to change the meas ured value memory status for user 5 press the SET button until the number 5 appea...

Page 12: ...Off button Otherwise continue making settings by pressing the SET button 2 8 Clearing the measured value memory The operation for clearing the measured value memory is the last item in the settings m...

Page 13: ...re for your blood glucose measurement as follows Step 1 Wash your hands thoroughly with warm water and soap Step 1 With the meter switched off insert the code chip included in the test strip package i...

Page 14: ...ically Step 5 The user identity number of the measured value memory where the result of the last blood glucose measurement was stored blinks on the display see Section 3 5 Use by multiple persons Step...

Page 15: ...ng the On Off button without inserting a test strip in the test strip receptacle Steps 3 6 are car ried out the same way as described before As opposed to the rst way with the inserted test strip the...

Page 16: ...lease button Step 9 Carefully press on the nger pad until a blood drop is formed 3 4 Carrying out the blood glucose measurement Step 1 Hold the drop of blood next to the right side of the test strip A...

Page 17: ...eeps The measured blood glucose value is displayed with the date and time of the measurement This data is stored in the measured value memory of the user Attention Hyperglycaemia warning blood glucose...

Page 18: ...t strip out of the meter after the measurement Properly dispose of used test strips and lan cets immediately after measurement so that they cannot injure infect anyone 3 5 Use by multiple persons When...

Page 19: ...the result of the last blood glucose measurement was stored see Section 3 5 Use by multiple persons Step 3 In order to look at the contents of the measured value memory of a certain user press the ME...

Page 20: ...le measurement Step 8 To exit memory mode switch the device off again with the On Off button Once the end of the measured value memory has been reached or if there are no other measured values availab...

Page 21: ...uring ranges can be checked with the different control solutions Normal blood glucose levels are simulated with the gluco comfort control solution GCS105 Low blood glucose levels are simulated with th...

Page 22: ...for protecting and storing the supplies 5 7 Disposing of the meter When making blood glucose measurements blood can get onto and into the meter There fore a danger of infection is posed by used meters...

Page 23: ...he device with another code chip or have the device checked by the manufacturer Error message Causes Measures Remarks 4 error Coding error in the device Remove the code chip and insert again If this e...

Page 24: ...nt After restarting the device insert the test strip into the device Error message Causes Measures Remarks 16 error Blood glucose level too high out of measuring range of gluco comfort If the measured...

Page 25: ...Housing is grey aluminium buttons and battery compartment cover are black BGM205 Housing is slate grey buttons and battery compartment are light blue Automatic switch off function 4 minutes when test...

Page 26: ...GCS205 tar get value about 40 mg dL gluco comfort high control solution GCS305 target value about 400 mg dL Transportation and storage conditions 4 C to 30 C at a relative humidity of 10 to 95 When u...

Page 27: ...tion are in vitro diagnostic supplies for personal use The CE marking of this product con rms that the relevant require ments of the EC directive 98 79 EC for these products are ful lled The blood glu...

Page 28: ...54 55...

Reviews: