46
47
Error
message
Causes
Measures
Remarks
8 error
Bad operating conditions:
The ambient temperature is
too low.
Adapt the ambient tem-
perature to the required
operating conditions (see
technical data).
10 error
No test strip was inserted
after starting the device.
Repeat measurement.
Due to a timeout error, the
device interrupted the measu-
rement.
11 error
No blood or control solution
was applied to the test strip.
Repeat measurement.
Due to a timeout error, the
device interrupted the measu-
rement.
14 error
Test strip must be removed.
Remove the test strip from
the device.
Due to a timeout error, the
device interrupted the measu-
rement.
15 error
After the device was re-
started, the code chip was
inserted in the device instead
of the test strip.
Repeat measurement.
After restarting the device, insert
the test strip into the device.
Error
message
Causes
Measures
Remarks
16 error
Blood glucose level too high
(out of measuring range of
gluco-comfort
).
If the measured value
does not correspond with
the way you feel, repeat
measurement.
If this error message appears
repeatedly, have the device
checked by the manufacturer.
17 error
Blood glucose level too low
(out of measuring range of
gluco-comfort
).
If the measured value
does not correspond with
the way you feel, repeat
measurement.
If this error message appears
repeatedly, have the device
checked by the manufacturer.
19 error
(BGM 105)
No user has yet been created
with the Health Manager.
Configure the device with the
Health Manager and enable at
least one user.
21 error
(BGM 105)
No wireless module found.
Switch off the device and
restart.
If this error message appears
repeatedly, have the device
checked by the manufacturer.