Attention:

If the levels are not reached, please repeat 

your measurements. If the values are still out-

side the specified tolerances, please contact 

the manufacturer of the blood glucose meas-

uring system. Do not use the blood glucose 

measuring system until it has been made sure 

that the displayed values are correct.

Note:

Use only 

gluco-comfort

 control solutions 

(GCS105, GCS205 or GCS305), since only 

these are adapted to the measuring principle 

of the Biocomfort blood glucose meter and 

gluco-comfort

 test strip combination.

 

Attention:

If the measuring cell of the test strip has 

direct contact with the tip of the control 

solution bottle, the enzyme of the measuring 

cell could get into the control solution, 

rendering it useless.

To keep the control measurement from being 

included in the average calculation of the 

blood glucose values, you must mark the 

control measurement as such. 

To do this, press the MEM button when the 

meter is ready for measurement (Step 1 in 

Section 3.4). In the display, the abbreviation 

“ctl” (control) appears to the right of the drop 

symbol. The changeover to control mode is 

confirmed with a beep. The function of the 

individual measuring ranges can be checked 

with the different control solutions:

−  Normal blood glucose levels are simulated 

with the 

gluco-comfort

 control solution 

GCS105.

−  Low blood glucose levels are simulated 

with the 

gluco-comfort

 control solution 

GCS205.

−  High blood glucose levels are simulated 

with the 

gluco-comfort

 control solution 

GCS305.

Notes:

−  Please follow the separate user instructions 

for the control solutions.

−  Shake the control solution before use.

−  The control solution may not be used past 

the expiration date. 

−  Store the control solution at room tempera-

ture.

−  The control range printed on the test  

strip packaging only applies for the  

control solution. It is not a recommended 

range for your blood glucose value!

−  The control solution is not a cleaning  

solution for the meter.

5.4 Evaluating the control measurement

When the measuring system is working  

correctly, the measured blood glucose  

values must correspond with the ranges 

which are listed on the test strip packaging  

for normal, low and high.

gluco-comfort��

Kontroll-Lösung�GCS�101

12ml

Normal���

zur�Eigenanwend��g

•�vor�der�Verwendung�gut�s��ütteln

•�nicht�zum�Verzehr�geeign��t

•�nicht�verbrauchte�Kontroll�ng�ist�30�Tage

���nach�Anbruch�der�Ampul���verwerfen:

•�Anbruchdatum:
Biocomfort�Diagnostics�Gm��&

Co.�KG

Bernhäuser�Straße�17

D�73765�Neuhausen�a.�d.�Fil�

gluco-comfort

mem

biocomfort

ctl

mg/dL

mem

Fu

nc

tion c

he

ck a

nd c

ar

5

Summary of Contents for BGM 105

Page 1: ...gluco comfort Blood glucose measuring systems BGM105 BGM 205 Operating manual Issued in 11 2006 REF 590107...

Page 2: ...the blood glucose measurement 28 3 4 Carrying out the blood glucose measurement 31 3 5 Use by multiple persons 35 4 Measured value memory 36 4 1 Storing the measured values 36 4 2 Calling the stored m...

Page 3: ...attery blinking dead battery 5 Number eld for Year Time User activated not activated Wireless connection activated not activated 6 Symbol for active wireless connection BGM 105 7 Code no indication wh...

Page 4: ...is operating manual contains important information for you It is mandatory that the safety measures mentioned in this operating manual be taken The Biocomfort blood glucose measure ment system may onl...

Page 5: ...ons for the test strips The test strips could be damaged by effects of temperature and light For this reason they should be used immediately after being removed from the test strip con tainer The test...

Page 6: ...tion regarding the control solution Please observe the separate user instructions for the control solution Always shake the control solution well before use The control solution can be damaged if expo...

Page 7: ...e Step 6 Afterwards check whether the battery has been correctlyinsertedbybrie ypressingtheOn Off button Inthiscase allelementsofthedisplayap pear for about two seconds as a function check Note Please...

Page 8: ...lue you desire Now you can go to the next parameter by pressing the SET button or you can exit the settings menu by pressing the On Off button Note If you keep the MEM button pressed you can change th...

Page 9: ...acti vate the minute display Step 10 Keep pressing the MEM button until the cor rect minute is displayed Step 11 If you are nished making settings you can exit the settings menu by pressing the On Of...

Page 10: ...1 Con guration of the model variant BGM205 In its delivered condition the measured value memory is only enabled for one user User 1 The con guration of the measured value memory can be found after th...

Page 11: ...ou wish you can set the status for user 2 by pressing the SET button Remark If for example you want to change the meas ured value memory status for user 5 press the SET button until the number 5 appea...

Page 12: ...Off button Otherwise continue making settings by pressing the SET button 2 8 Clearing the measured value memory The operation for clearing the measured value memory is the last item in the settings m...

Page 13: ...re for your blood glucose measurement as follows Step 1 Wash your hands thoroughly with warm water and soap Step 1 With the meter switched off insert the code chip included in the test strip package i...

Page 14: ...ically Step 5 The user identity number of the measured value memory where the result of the last blood glucose measurement was stored blinks on the display see Section 3 5 Use by multiple persons Step...

Page 15: ...ng the On Off button without inserting a test strip in the test strip receptacle Steps 3 6 are car ried out the same way as described before As opposed to the rst way with the inserted test strip the...

Page 16: ...lease button Step 9 Carefully press on the nger pad until a blood drop is formed 3 4 Carrying out the blood glucose measurement Step 1 Hold the drop of blood next to the right side of the test strip A...

Page 17: ...eeps The measured blood glucose value is displayed with the date and time of the measurement This data is stored in the measured value memory of the user Attention Hyperglycaemia warning blood glucose...

Page 18: ...t strip out of the meter after the measurement Properly dispose of used test strips and lan cets immediately after measurement so that they cannot injure infect anyone 3 5 Use by multiple persons When...

Page 19: ...the result of the last blood glucose measurement was stored see Section 3 5 Use by multiple persons Step 3 In order to look at the contents of the measured value memory of a certain user press the ME...

Page 20: ...le measurement Step 8 To exit memory mode switch the device off again with the On Off button Once the end of the measured value memory has been reached or if there are no other measured values availab...

Page 21: ...uring ranges can be checked with the different control solutions Normal blood glucose levels are simulated with the gluco comfort control solution GCS105 Low blood glucose levels are simulated with th...

Page 22: ...for protecting and storing the supplies 5 7 Disposing of the meter When making blood glucose measurements blood can get onto and into the meter There fore a danger of infection is posed by used meters...

Page 23: ...he device with another code chip or have the device checked by the manufacturer Error message Causes Measures Remarks 4 error Coding error in the device Remove the code chip and insert again If this e...

Page 24: ...nt After restarting the device insert the test strip into the device Error message Causes Measures Remarks 16 error Blood glucose level too high out of measuring range of gluco comfort If the measured...

Page 25: ...Housing is grey aluminium buttons and battery compartment cover are black BGM205 Housing is slate grey buttons and battery compartment are light blue Automatic switch off function 4 minutes when test...

Page 26: ...GCS205 tar get value about 40 mg dL gluco comfort high control solution GCS305 target value about 400 mg dL Transportation and storage conditions 4 C to 30 C at a relative humidity of 10 to 95 When u...

Page 27: ...tion are in vitro diagnostic supplies for personal use The CE marking of this product con rms that the relevant require ments of the EC directive 98 79 EC for these products are ful lled The blood glu...

Page 28: ...54 55...

Reviews: