Using Global Logger Software
Pulse Channel Calibration
To change pulse channel settings, refer to the
Pulse Input
fields in the
Global Logger Setup
window. The
pulse channel records a predetermined value each time the sensor detects a pulse signal. This feature is of-
ten used to track rainfall levels.
To change the displayed unit type (i.e. meters, feet, etc.), tap the
Sensor Name or EU
(engineering units)
field and enter the unit type you are calibrating to. Note that the
Sensor Name or EU
field only serves as a
text label.
To set the recorded value for a single pulse reading, tap the
Value for Single Pulse
field and enter the de-
sired predetermined value.
To set the number of decimal place values to display, tap the
Decimal Places
field and enter the number of
decimal places (0 – 6), or click the up/down arrows to the right of the text field to increase or decrease the
value by 1. After changing settings, click the
OK
button. To cancel changes, click
Cancel
.
Verify that
High EU
and
Low EU
fields reflect desired unit readings. Also, verify that the
High Raw
field
reflects a range of 60,000 – 65,000 and that the
Low Raw
field reflects a range of 10,000 – 15,000.
Adjust the
Decimal Places
field according to the accuracy percentage rating. The number of decimal
places should not exceed the accuracy of the sensor. After calibrating, click the
OK
button. To cancel
changes, click
Cancel
.