![Global Water GL400 User Manual Download Page 26](http://html1.mh-extra.com/html/global-water/gl400/gl400_user-manual_2233390026.webp)
Page 26
Global Water
●
(800)876-1172
●
www.globalw.com
For both environmental and lab calibrations,
wait at least five seconds for the sensor to
adjust to current sensing levels. Click the
Next
button.
Enter
0
in the
lowest value
text field and
click the
Next
button.
For lab calibrations, release the sensor from calibration meter. For environmental calibrations, take sensor
out of targeted environment.
For both environmental and lab calibrations,
wait at least five seconds for the sensor to
adjust to 0 sensing levels. Click the
Finish
button.
*See
Appendix B: Worksheets
, pages 36
Note, that if performing an envi-
ronmental calibration at a frac-
tion of total sensor range, the
bar graphs in the main
Global
Logger
window will interpret
full scale range at maximum
calibration readings instead of
actual total sensor range. All
other logger operations will per-
form correctly. To revise cali-
bration settings so that the
graphs will function correctly,
follow the worksheet
Calibrat-
ing the Bar Graph to Higher
or Lower Range
in
Appendix
B
*.
Chapter 3