USER MANUAL

 

DIGITAL READOUT  VISION 518

 

MT02_A31_B_VI518_GIVI_ENG      rev. A

 

Pag. 26/32

 

 

 

Examples of programming with the axis in angular reading: 

 

NC

NC

NO = normally open

dt

NO

NC = normally closed

NO

NO

Example  1:   Relay 1 NO at the angle 90°  with a sector angle of 2°

91°

90°

89°

Example  2:   Relay 1 NC at the angle 90° with a sector angle of 2°

91°

90°

89°

Example  3:   Relay 1 NO at the angle 0° with a sector angle of 10°

355°

 

 

 

Summary of Contents for VISION 518

Page 1: ...User Manual Digital Readout V VI IS SI IO ON N 5 51 18 8 www givimisure it...

Page 2: ...ON F 11 p 13 MIDPOINT CALCULATION F 15 p 13 INVERSION OF COUNTING DIRECTION F 22 p 14 LOAD POSITION PRESET F 23 p 14 DISABLING THE AUTOMATIC REF SEARCH F 24 p 15 COUNTING MODE F 25 p 16 LINEAR CORRECT...

Page 3: ...ith a number of specific functions for any type of machine tool PRELIMINARY REMARKS Disposal of waste electrical and electronic equipment WEEE European Council Directive 2002 96 EC The use of the WEEE...

Page 4: ...trument is connected must be equipped with a sectioning device in compliance with the regulations in force positioned close to the instrument PREVENTION To avoid fires or explosions the instrument sho...

Page 5: ...USER MANUAL DIGITAL READOUT VISION 518 MT02_A31_B_VI518_GIVI_ENG rev A Pag 5 32 DIMENSIONAL SPECIFICATIONS DRILLING TEMPLATE 92 x 45 mm 140 96 48 10 48 96 10 140 48...

Page 6: ...1 Relay 1 common contact NA1 N O Relay 1 contact C2 Relay 2 common contact NA2 N O Relay 2 contact V Encoder power supply output 12 V 05 V 0 V 0 V Encoder power supply LOAD Load position input connect...

Page 7: ...ides a series of visual signals to support the user during configuration and use When pressing a key messages appear on the display according to the operation function see this manual Finally the inst...

Page 8: ...ot consider shifts of the carriage made whenever the mains power supply is disconnected due to inertia manual maneuvers or thermal expansion Whenever any of the above occurs the position shown in the...

Page 9: ...rruption of power supply voluntary or not occurs the instrument proposes to the operator the zero reference REF search to avoid the displaying of incorrect positions LAST POSITION This is the case for...

Page 10: ...the axis has terminated the zero reference REF search displaying an error message at the pressure of any key except CLR The operator can then decide to A Complete the zero reference search B Cancel i...

Page 11: ...d only affects the counter of the selected counting system For instance it is possible to reset the ABS counter in a given point ORIGIN and operate in INCREMENTAL counting that is handier and more ver...

Page 12: ...al point therefore from 999999 9 to 999999 9 for 100 m resolution from 99999 99 to 99999 99 for 10 m resolution from 9999 999 to 9999 999 for 1 m resolution The F 00 function can be used to delete the...

Page 13: ...he carriage along the axis Let us suppose that in that position the axis display will show the value 30 00 which is not necessary to reset Press 1 5 3 0 0 0 Note The decimal point of the position will...

Page 14: ...the operator that the inversion of direction has been executed Press because the axis has to be inverted d i r Press the selection is confirmed and the axis is inverted X X X X X X X X After having c...

Page 15: ...e described below Press the first digit will flash X X X X X X X Press and the value will be set preset 5 0 0 0 It is possible to recall the load position also with an external contact connected to th...

Page 16: ...The counting mode can be direct double quadruple or mono directional Press 2 5 C n t n 1 The display will show the direct counting mode n1 Press to select the double counting mode n2 C n t n 2 Press t...

Page 17: ...performed moving the axis by 400 00 mm value shown on the display but an error has been detected positive or negative For instance the measured piece is either longer or shorter than intended That me...

Page 18: ...irm the selection After having confirmed with the EXE key the display will be enabled for counting While choosing the reading mode consider that A RADIAL READING the display shows the same value of th...

Page 19: ...d resolution r 0 0 Press to confirm the selection 1 At any pressure of the key the decimal point shifts to the lower resolution 2 The variable resolution set will be maintained also after turning off...

Page 20: ...olution is 3 6 seconds of an arc 0 001 that can be obtained with an encoder having 90 000 PPR If the instrument has been configured by the Manufacturer in angular reading the operator can select the r...

Page 21: ...ted only if the instrument has been required with this option Vice versa the operator should contact the Manufacturer To activate the function Press 4 0 r E L E no Press to modify the setting r E L E...

Page 22: ...it and press EXE 0 0 0 0 0 0 5 To set a time of 2 sec for relay 2 press 4 1 2 0 0 Note The key is used to select the desired relay 2 in our example The program has been completed and in our example th...

Page 23: ...tation direction To disable the timed output and return to the standard relay output it is sufficient to set a timing duration of 0 seconds on the concerned relay Axis in linear reading Let us suppose...

Page 24: ...relays activate according to the formula intervention angle circular sector Let us suppose that the operator wants to activate the relay 1 at the position of 90 with a circular sector having an angle...

Page 25: ...is in linear reading NC NC NO normally open dt NO NC normally closed NO NO Example 1 Relay 1 NO at the position 100 Relay 2 NC at the position 100 RELAY 2 RELAY 1 100 100 0 Example 2 Relay 1 NO at the...

Page 26: ...ng with the axis in angular reading NC NC NO normally open dt NO NC normally closed NO NO Example 1 Relay 1 NO at the angle 90 with a sector angle of 2 91 90 89 Example 2 Relay 1 NC at the angle 90 wi...

Page 27: ...8 9 n c RX TX n c 0 V n c n c n c n c The SLAVE needs to have the same parameters If the digital readout has the serial output each pressure of the key will start the transmission Data have the follo...

Page 28: ...on of displayed positions on the serial output Press 5 5 S E n d no Press to select S E n d yes Press to confirm the selection Data on ABS or INC positions are sent every 0 4 sec in ASCII mode and hav...

Page 29: ...call a function that cannot be selected with the key Press to enter the function selection F F n Press F 0 0 0 0 Use and to enter the number of the desired function and press EXE F 9 8 7 7 1 Press to...

Page 30: ...24 Vac Memory permanent for configuration and user settings last data operating memory Linear resolution 200 100 50 20 10 5 2 1 0 5 m 0 01 0 005 0 002 0 001 0 0005 0 0002 0 0001 0 00005 0 00002 inch...

Page 31: ...lacement of those parts that have been damaged by negligence or misuse improper installation or maintenance maintenance performed by unauthorized personnel transport or any other circumstance that exc...

Page 32: ...IGHE OTTICHE OPTICAL SCALES SISTEMI MAGNETICI MAGNETIC SYSTEMS ENCODER ROTATIVI ROTARY ENCODERS VISUALIZZATORI DIGITAL READOUTS POSIZIONATORI POSITION CONTROLLERS GIVI MISURE S r l A SOCIO UNICO Via A...

Reviews: