GIMA spa
DIATERMO MB240 – DIATERMO MB380
Manuale d’ Istruzioni / Instruction’s Manual
MA129IGBg
43
MONOPOLAR COAG 3 (SPRAY):
80 W +/- 10%
BIPOLAR
CUT:
90
W
+/-
10%
BIPOLAR
COAG:
80
W
+/-
10%
WARNING: WHEN THE DUAL OPERATIVE MODE IS SELECTED, THE SUM OF THE OUTPUT OF BOTH THE
MONOPOLAR AND BIPOLAR CURRENT MUST NOT EXCEED 400 WATTS.
If the output power presetting is made
after DUAL operative mode is selected it will not be possible to increase the total power up the previously said value of 400
Watts. In the case the output power is preset before of the DUAL operative mode is selected, and the sum of the preset power
exceeds the 400 Watts value, when the DUAL operative mode is selected, the digital displays showing the preset power, will
blink and the output will be interdicted untill allowed preset condition had been restored.
Reference Load Resistance:
MONOPOLAR
CUT:
300
Ω
MONOPOLAR CUT/COAG 1:
300
Ω
MONOPOLAR CUT/COAG 2:
300
Ω
MONOPOLAR
COAG
1:
300
Ω
MONOPOLAR
COAG
2:
300
Ω
MONOPOLAR COAG 3 (SPRAY):
2 k
Ω
BIPOLAR
CUT:
100
Ω
BIPOLAR
COAG:
100
Ω
No Load Max Voltage:
MONOPOLAR
CUT:
1.4
kVpp
+/-10%
MONOPOLAR CUT/COAG 1:
1.4 kVpp +/-10%
MONOPOLAR CUT/COAG 2:
1.4 kVpp +/-10%
MONOPOLAR
COAG
1:
1.4
kVpp
+/-10%
MONOPOLAR
COAG
2:
1.4
kVpp
+/-10%
MONOPOLAR COAG 3 (SPRAY):
4.2 kVpp +/-10%
BIPOLAR
CUT:
0.5
kVpp
+/-10%
BIPOLAR
COAG:
0.5
kVpp
+/-10%
Crest Factor:
MONOPOLAR
CUT:
1.4
MONOPOLAR CUT/COAG 1:
1.5
MONOPOLAR CUT/COAG 2:
1.7
MONOPOLAR
COAG
1:
2.3
MONOPOLAR
COAG
2:
3.5
MONOPOLAR COAG 3 (SPRAY):
7
BIPOLAR
CUT:
1.4
BIPOLAR
COAG:
1.5
Output Frequency:
MONOPOLAR: 475 kHz +/- 10%
BIPOLAR:
525 kHz +/- 10%
Modulation Frequency:
MONOPOLAR
CUT:
NONE
MONOPOLAR CUT/COAG 1:
10 kHz
MONOPOLAR CUT/COAG 2:
10 kHz
MONOPOLAR COAG 1:
10
kHz
MONOPOLAR COAG 2:
10
kHz
MONOPOLAR COAG 3 (SPRAY):
10 kHz
BIPOLAR
CUT:
NONE
BIPOLAR
COAG:
10
kHz
Allowed output time:
DUTY-CYCLE: 10 seconds output-25 seconds pause
Allowed maximum total output power: 400W.
5.2 OUTPUT POWER DIAGRAMS
See pages from 22 up to 25.