GIMA spa
DIATERMO MB240 – DIATERMO MB380
Manuale d’ Istruzioni / Instruction’s Manual
MA129IGBg
33
3. INSTALLATION
Inspect the equipment for damage during transit. Any damage should be reported to the carrier immediately.
Unpack the equipment and carefully study the documentation and operating instruction supplied.
The mains input voltage, indicated up the equipment mains input, must agree with the local mains voltage (mains
frequency: 50-60 Hz).
).
The predisposition of the correct mains voltage is performed in the following way:
(A-B) Extract the fuse holder drawer from the power module.
(C) Insert the fuses making reference to the following chart:
Mains Voltage 110-120 V
Delayed fuse 10AT - 5 x 20 mm
Mains Voltage 220-240 V
Delayed fuse 5AT - 5 x 20 mm
(D) Extract and rotate the detachable part in way to read the correct voltage in the (E) window – reinsert the fuse
holder in the module.
Connect the mains cable to a mains outlet that has good earth connection.
OPERATION OF THE EQUIPMENT WITHOUT EARTH CONNECTION IS FORBIDDEN
Position the equipment such that there is sufficient room at the ventilation slots at the rear of the unit. The ventilation
slots in the rear of the equipment must be kept unobstructed. Air circulation must not be restricted in any way.
Operate the equipment only in dry surroundings. Any condensation that occurs must be allowed to evaporate before
putting the equipment into operation. Do not exceed the permissible ambient temperature or humidity.
Before using the unit, it is necessary connect the cable to the patient plate. Both when single plate electrodes and when
split plate electrodes are used it is necessary to confirm the impedance acceptance by pressing the key OK (see
paragraph 4.3.5). In this way, if the value of the impedance is acceptable, the OC indicator light will stop flashing and
the alarm to sound.
Environmental conditions:
Work
Transit/storage
Temperature:
from 10 °C up to 40 °C
from -10 °C up to 50 °C
Relative moisture:
from 30% up to 75%
from 10 % up to 100%
Atmospheric pressure:
from 70 kPa up to 106 kPa
from 50 kPa up to 106 kPa