31
ENGLISH
Table 2 - Guidance and manufacturer’s declaration- electromagnetic immu-
nity for all EQUIPMENTAND SYSTEMS
This device is intended for use in the electromagnetic environment specified
below. The customer or the user of the equipment or system should assure that
it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge
(ESD)
IEC61000- 4-2
±8 kV contact
±15kV air
±8 kV contact
±15kV air
Floors should be wood,
concrete or ceramic tile.
if floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
If ESD interfere with the
operation of equipment,
counter measurements
such as wrist strap,
grounding shall be
considered.
Electrical fast
transient/bur st
IEC61000- 4-4
±2kV for power
Supply lines
±1 kV for input/
output lines
±2kV for
power Supply
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Sovratensione
IEC 61000-4-5
±1kV line (s)
to line(s)
±2kV line(s)
to earth
±1kV
differential
mode
±2kV
common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.