in the gas loop is evacuated and its concentration measured is decreased and
reaches zero, when exhaling, CO2 enters the breathing loop and its concentration
rises rapidly and is kept at a certain platform, at the end tidal breathing it reaches
maximum. In this repeated way, a real-time and high or low waveform is
formed and by the virtue of this waveform, the device calculates the respiration
status and also by me asuring respiration cycle, the device meantime calculates
the respiration rate.
Update/Averaging Time: Option of every breath
Warm Up Time: <15 seconds
Total system response time: < 1s
Memory: 24 hours on Screen Trend and Numeric
Respiration Rate
Range: 3 - 150 rpm
Accuracy
: ±1% of reading or ±1 rpm whichever is greater
Memory
: 24 hours on Screen Trend and Numeric
Sensor
: Adapter for intubated Patients
Alarm limit
The high alarm limits of EtCO2: 22-99mmHg
The low alarm limits of EtCO2: 10-60mmHg
The high alarm limits of respiration rate: 5-60 breaths/min
The low alarm limits of respiration rate: 4-40 breaths/min
Power
AC Input: 100V - 240V, 50Hz/60 Hz to 5VDC Adapter with 5V mini USB adapter
Cable.
Battery
Type: Built-in rechargeable lithium battery pack (3.7V, 1200mAH)
Charging Time: 4 hours from flat
Operating Time: 6 hours on full charge
Operating Conditions
Temperature: 5 to 40 °C
Humidity: 30%~75%