13
ENGLISH
Chapter 3 - Introduction of Monitor
Figure 1
(1) Screen: Displays waves, menu, alarm and all measuring parameters.
(2) S button: Press this button to move the cursor when menu is activated.
(3)
+/
Multifunction button
a) When menu is activated, to press this button as confirmation button or increase
(+) button.
b) In the main screen, if the bluetooth printer is equipped, to press this button for
two seconds to drive the printer to print CO2 reaction curve and result data of all
parameters.
(4) Multifunction button
a) When menu is activated, to press this button to decrease data selected.
b) In the main display screen, to press this button to silence alarm for two minutes.
c) In the main display screen, to press this button over 2 seconds, the display
screen will change to big font display mode as the figure showing:
Summary of Contents for 33831
Page 36: ......