Alarm sound:
Alarm sounds as following description. Time interval cannot be changed.
Level of alarm
Sound
Sound
pressure
Medium priority alarm
Beep-Beep Beep”, triggered
each 8 seconds
45-70dB
Alarm light:
Alarm light looks like following description
Level of alarm
Light
Intermediat alarm
Parameter indicator turn yellow, blinking with
frequency of 0.5Hz
Alarm silence:
In the main display screen (menu is not open), to press the button
to silence
alarm for two minutes and meantime, trumpet icon becomes , two minutes later,
trumpet icon becomes
, meanwhile, alarm begins to
work if there is sound alarm. If to press the button
in this period, the silence
alarm can be released. Silence alarm time will not vary with operator. When silence
alarm is on, physical alarm and technical alarm both will be silent.
Alarm counterplan:
WARNING
: Always check status of patient as alarm is triggered.
Check the alarm information displayed on the screen, correctly identify the alarm,
and reasonably handle the alarm according to the cause of the alarm.
Check patient’s status.
Identify type or parameter of alarming.
Find the reason.
Turn off alarming if necessary.
Check alarm after removing alarming condition.