IMMUNITY
test
IEC 60601
test level
Complia
nce level
Electromagnetic
environment - guidance
RF condotta
RF IEC
61000-4-6
RF irradiata
IEC
61000-4-3
3 Vrms da 150
kHz a 80MHz
(6V in ISM e
bande radio
amatoriali tra
0,15MHz e 80
MHz)
10 V/m
da 80 MHz a
2,7 GHz
3 Vrms
da 150 kHz
a 80MHz (6V
in ISM
e bande
radio
amatoriali tra
0,15MHz
e 80 MHz)
3 V/m
Portable and mobile RF
communications equipment
should be used no closer
to any part of this device,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
= 1,2√P
= 1,2√P 80MHz a 800MHZ
= 2.3√P 800 2,7
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
,a should be less than the
compliance level in each
frequency range Interference
may occur in the vicinity of
equipment marked with the
following symbol.