30
ENGLISH
Guidance and manufacture’s declaration – electromagnetic immunity
The Non Contact Infrared Body Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user ofthe Non Contact Infrared Body
Thermometer should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
10V/m
80 MHz
to 2.5 GHz
Not
applicable
10 V/m
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the Non Contact Infrared Body
Thermometer, including cables,
than the recommended sepa-
ration distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation
distance
80MHz to 800MHz
80MHz to 800MHz
Where
P
is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufacturer
and d is the recommended sep-
aration distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment
marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
d
= 1.2
√P
d
= 1.2
√P
d
= 2.3
√P