Environmental
Operating Condition
Operating Temperature: 15℃~40℃
(59°F ~104°F),
Relative Humidity ≤85%,
atmospheric pressure:70 Kpa -106 Kpa
Transport and Storage
Condition
Storage Temperature:-20°C -55°C /
(-4°F - 131°F),
Relative Humidity ≤93%,
Atmospheric pressure:70 Kpa -106 Kpa
Dimension and Weighting
Weight (without batteries)
80g
Size
L:122mm X W:96mm X H:42mm
Compliance
Item
Compliant with
Equipment classification
Safety Standards: EN 60601-1: 2006+A1:2013,
EN 60601-1-2: 2015
Type of protection
Internally powered equipment (on battery power)
Degree of protection
Non Applied part
Front panel and case
labeling
EN ISO15223-1:2012
Temperature
EN ISO80601-2-56:2017
Home healthcare
environment
EN 60601-1-11:2015
Protection against electric shock
Internally powered ME equipment
Applied part
Non Applied part
Protection against harmful ingress of
water or particulate matter
IP22
Mode of operation
Continuous operation
Safety classification of ME EQUIPMENT
Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT