21
Technical Specifications
Measurement Unit
℃/℉
Operating mode
Adjusted mode (Body mode)
Direct mode (surface mode)
Reference Body Site
Axillary
Rated output range
Body mode: 34.0℃~43.0℃ / 93.2°F -109.4°F
Surface mode: 0℃~100℃ / 32℉~212℉
Output Range
Body mode: 34.0℃~43.0℃/ 93.2°F -109.4°F
Surface mode: 0℃~100.0℃/ 32°F -212°F
Laboratory Accuracy
Body mode:
34.0℃~34.9℃:±0.3℃ / 93.2°F-94.8°F: ±0.5°F;
35.0℃~42.0℃:±0.2℃ / 95.0°F-107.6°F: ±0.4°F;
42.1℃~43.0℃:±0.3℃ / 107.8°F-109.4°F: ±0.5°F;
Surface mode: ±2℃/±3.6°F.
Display Resolution
0.1℃/0.1℉
Three-color Backlight
(Color Alarm)
35.5-37.3°C / 95.9-99.1°F:
Green (Normal Temperature);
37.4-38.0°C (Alarm point) / 99.3-100.4°F:
Yellow (Slight Fever)
38.1-43.0°C / 100.6-109.4°F:
Red (High Fever)
Note:
1.Surface mode is always with Green backlight.
2. In body mode 34.0-35.4°C is with green backlight.
Auto Power Off Time
≤18s
Measuring Time
≤2s
Measuring Distance
1 cm -5cm (0.4 in -2in)
Memories
50
Power Supply
Requirements
Batteries
1.5V (AAA) Alkaline battery x2 (IEC Type LR03)
Adaptable Range
2.6V~3.6V
ENGLISH