![background image](http://html1.mh-extra.com/html/gima/25552/25552_manual_2228192006.webp)
ORAL USE (FIGURE 1)
· The mouth should remain closed 3 to 5 minutes before attempting a reading.
· Place the thermometer in the mouth under the tongue so that it rests to the left or right
of the root of the tongue.
· Use downward tongue pressure to hold the thermometer in place.
· Hold the thermometer to keep it from sliding around in the mouth.
· When the reading is ready, the thermometer will emit a beep sound, remove
the thermometer and read the result.
Figure 1
UNDERRARM USE (FIGURE 2)
· The underarm should remain clamped 3 to 5 minutes before attempting a reading.
· Place the probe tip in the center of the underarm.
·
·
Lock the probe tip under the arm, using the arm to slightly apply pressure inwards.
When the reading is ready, the thermometer will emit a beep sound, remove
the thermometer and read the result.
CARE AND MAINTENANCE
· Wipe the thermometer with a soft clean cloth before and after each use.
· Never put the thermometer into any liquid.
· Do not shock, knock or move the thermometer when using it.
· The thermometer is not a toy; please keep it out of the reach of children.
· Adult supervision is required when taking a temperature on children.
· Contact your physician for specific information about your temperature readings.
·
·
Store the thermometer in a clean, safe and dry location. Do not store the thermometer where it will be exposed
to direct sunlight, dust, etc. Keep it far from heat source. Storage temperature: -10°C~+60°C; Storage humidity <95%
Do not use a mobile phone near the thermometer or in places where strong static electricity or electromagnetic field
at present.
SPECIFICATIONS
Measurement Range
32.0°C~43.9°C (90.0°F~111.0°F)
“Lo”displayed when the thermometer of the sensing section is less than 32.0°C (90.0°F)
“Hi” displayed when the thermometer of the sensing section is greater than 43.9°C (111.0°F)
Measurement Accuracy
±0.1°C (35.0°C~41.0°C)
±0.2°F (95.0°F~105.8°F)
Temperature Display
3.5-digit LCD display
Resolution
0.1°C (0.1°F)
Buzzer Notification
Press ON/OFF button or the temperature measurement is complete; the buzzer will emit
beep sound.
Memory Function
The last reading will automatically be stored in memory.
Usage Environment
Surrounding Temperature: 10°C~40°C (50°F~104°F), Relative Humidity: 30%RH~85%RH
In the 16°C~31°C environment temperature range, measurement can be done in 8 to 10
seconds. Otherwise it will need 20 seconds to 2 minutes to complete temperature measu-
rement.
Storage Environment
Surrounding Temperature: -10°C~+60°C (14°F~140°F), Relative Humidity: 10%RH~95%RH
Battery
LR41
Reference Standard
EN12470-3: 2000; clinical thermometers ASTM E-1112; electronic thermometers
Figure 2
ENGLISH