18
Piston assembly and seals must be degreased with
i
sopropanol or
e
thanol before being immersed in de-
contamination solution in separate vessel.
Remove the tip-ejector, the tip-holder and the con-
necting nut.
Immerse them in the cleaning solution.
Degrease the piston assembly, the seals and immerse
them in another vessel.
Rinse each component with distilled water.
Leave the parts to dry by evaporation (or wipe with
a soft cloth the tip-ejector, the tip-holder and con-
necting nut).
Lubricate the piston assembly and the seals.
Reassemble the piston assembly, the tip-holder and
the tip-ejector.
13 - sPecIFIcatIons
PIPETMAN G is a high quality pipette that offers excel-
lent accuracy and precision. The figures given in the
“Gilson Maximum Permissible Errors” table (page 19)
were obtained using PIPETMAN DIAMOND Tips. These
figures are guaranteed only when genuine PIPETMAN
DIAMOND Tips are used.
Each pipette is inspected and validated by qualified
technicians in accordance with the Gilson Quality System.
Gilson declares that its manufactured pipettes comply
with the requirements of the ISO 8655 standard, by
type testing.
The adjustment is carried out under strictly defined and
monitored conditions (ISO 8655-6).
The data given in the tables conform to the ISO
8655-2 Standard.
With a precise pipetting technique (see Chapter
6 - General guidelines for good pipetting) the P2G
model may be used to aspirate volumes as low as
0.1 µL and the P10G model as low as 0.5 µL.