15
The use of other lubricant cancels the warranty
of this pipette.
Reassemble the pipette (refer to the figure, page 13).
Tighten the connecting nut (turn clockwise).
Refit the tip-ejector (see above).
changing the seals
The O-ring and seal are on the piston;
they must not be
autoclaved
, if worn or damaged in any way (chemical
or mechanical), they must be replaced. As the models
P2G and P10G contain miniaturized parts, it is best not
to disassemble these pipettes yourself, please contact
your local Gilson authorized Service Center.
The dimensions of the O-ring vary depending on the
model of pipette.
Remove the tip-ejector (see above).
Unscrew the connecting nut by turning it counter-
clockwise.
Carefully separate the lower and upper parts.
Remove the piston assembly, O-ring and seal.
If necessary clean the piston and replace the seal;
lubricate them lightly. Please place them in the
correct order.
Reassemble the pipette (refer to the figure, page 13).
Tighten the connecting nut (turn clockwise).
Refit the tip-ejector (see above).
12 - cLeanIng and decontamInatIon
PIPETMAN G is designed so that the parts normally in
contact with liquid contaminants, can easily be cleaned
and decontaminated. However, because the models
P2G and P10G contain miniaturized parts, it is best not
to disassemble these pipettes yourself; please contact
your local Gilson authorized Service Center.