19
gilson maximum Permissible errors
Maximum Permissible Errors
Model
Volume
Gilson
ISO 8655
(Reference)
(µL)
systematic
Random
systematic
Random
error (µL)
error (µL)
error (µL)
error (µL)
P2G
Min
0.2
± 0.024
0.012
± 0.08
0.04
(f144054P)
0.5
± 0.025
0.012
± 0.08
0.04
(f144054M)
Max.
2
± 0.030
0.014
± 0.08
0.04
P10G
Min.
1
± 0.025
0.012
± 0.12
0.08
(f144055P)
5
± 0.075
0.030
± 0.12
0.08
(f144055M)
Max
10
± 0.100
0.040
± 0.12
0.08
P20G
Min.
2
± 0.10
0.03
± 0.20
0.10
(f144056P)
10
± 0.10
0.05
± 0.20
0.10
(f144056M)
Max.
20
± 0.20
0.06
± 0.20
0.10
P100G
Min.
10
± 0.35
0.10
± 0.80
0.30
(f144057P)
50
± 0.40
0.12
± 0.80
0.30
(f144057M)
Max.
100
± 0.80
0.15
± 0.80
0.30
P200G
Min.
20
± 0.50
0.20
± 1.60
0.60
(f144058P)
100
± 0.80
0.25
± 1.60
0.60
(f144058M)
Max.
200
± 1.60
0.30
± 1.60
0.60
P1000G
Min.
100
± 3
0.6
± 8
3.0
(f144059P)
500
± 4
1.0
± 8
3.0
(f144059M)
Max.
1000
± 8
1.5
± 8
3.0
P5000G
Min.
500
± 12
3
± 40
15
(f144066)
2500
± 15
5
± 40
15
Max.
5000
± 30
8
± 40
15
P10mlG
Min.
1000
± 30
6
± 60
30
(f144067)
5000
± 40
10
± 60
30
Max. 10000
± 60
16
± 60
30
each pipette model (except P5000G and P10mlG) has two
different ordering references to identify the kind of tip-ejector
required. For a pipette with a plastic tip-ejector, the ordering
reference is ended by the letter P, for a pipette with a stainless
steel tip-ejector, the ordering reference is ended by the letter m.
ex: For a P10G model with the plastic tip-ejector the ordering
reference is F144055P. For the same pipette with a stainless steel
tip-ejector, the ordering reference is F044055m.
Summary of Contents for Pipetman P1000G