background image

3

the  PiPetman  Fixed  can  be  easily  re-adjusted  in 

the laboratory, with compensation for variations in 

fluid viscosity and density. a special calibration key 

is provided.

the  nominal  volume  of  each 

model is shown on the small label 

on the push-button.

PiPetman Fixed is in accordance 

with isO 8655 standard.

2 - SpEcIfIcatIonS

PiPetman Fixed is a high quality pipette that offers 

excellent accuracy and precision. the figures given 

in  the “Gilson  maximum  Permissible  errors”  table 

were  obtained  using “PiPetman  diamOnd tips”. 

these  figures  are  only  guaranteed  using  genuine 

PiPetman diamOnd tips. 

The data given in the table conform to the ISO 8655-2 

Standard.

 

 

Maximum Permissible Errors

Model 

Volume

 

Gilson 

ISO 8655

(

Reference) 

  (µL) 

Systematic  Random 

Systematic  Random

 

 

error (µL) 

error (µL) 

error (µL)  error (µL)

F2 

(F123770) 

± 0.08 

≤ 0.03  

± 0.08 

≤ 0.04

F5 

(F123771) 

± 0.10 

≤ 0.04  

± 0.125 

≤ 0.075

F10

 (F123772) 

10 

± 0.10 

≤ 0.05 

± 0.12 

≤ 0.08

F20

 (F123604) 

20 

± 0.20 

≤ 0.06 

± 0.2 

≤ 0.1

F25

 (F123775) 

25 

± 0.25 

≤ 0.07 

± 0.5 

≤ 0.2

F50

 (F123778) 

50 

± 0.40 

≤ 0.15 

± 0.5 

≤ 0.2

F100

 (F123784) 

100 

± 0.80 

≤ 0.25 

± 0.8 

≤ 0.3

F200

 (F123605) 

200 

± 1.60 

≤ 0.30 

± 1.6 

≤ 0.6

F250

 (F123787) 

250 

± 3.00 

≤ 0.75 

± 4.0 

≤ 1.5

F300

 (F123788) 

300 

± 3.50 

≤ 0.75 

± 4.0 

≤ 1.5

F400

 (F123789) 

400 

± 3.60 

≤ 0.80 

± 4.0 

≤ 1.5

F500

 (F123790) 

500 

± 4.00 

≤ 1.00 

± 4.0 

≤ 1.5

F1000

 (F123606) 

1000 

± 8.0 

≤ 1.3 

± 8.0 

≤ 3.0

F5000

 (F123607) 

5000 

± 30 

≤ 8 

± 40 

≤ 15.0

Summary of Contents for Pipetman Fixed

Page 1: ...User s Guide...

Page 2: ...sable tipsmeannoneedtohandle tips protecting the user from contamination Thetip holder theconnect ing nut and the tip ejector are autoclavable for 20 minutes at 121 C and 0 1 MPa pressure CONTENTS pag...

Page 3: ...y guaranteed using genuine PIPETMAN Diamond Tips The data given in the table conform to the ISO 8655 2 Standard Maximum Permissible Errors Model Volume Gilson ISO 8655 Reference L Systematic Random Sy...

Page 4: ...n measurements for each test volume 3 OPERATION Place a tip on the tip holder Press the tip on firmly using a slight twisting motion to ensure a positive airtight seal See Chapter 6 for the appropriat...

Page 5: ...calibrated using distilled waterandveryhighprecisionbalances Thenominal valueofthePipetmanFixedmaybeslightlyadjusted to compensate for liquids of different density or viscosity Adjustmentofthefactoryc...

Page 6: ...n displayrepresents0 4 LfortheF100 1 10of4 L thecalibra tiontoolmustbeturned3unitsuntilthedisplayreads B0 2 Itispossibletoalterthevolumebymorethanonefull turn in the counterclockwise direction In this...

Page 7: ...rate Pipetman Fixed therefore for optimum performance we recommend using PIPETMANDiamondTipswithyourPipetmanFixed PIPETMAN Diamond Tips have the Gilson logo engraved on their collar ensuring that you...

Page 8: ...oided Gilson ssterilizedDiamondFilterTipsarecertifiedfree ofdetectableRNase DNase DNA RNA andprotease pipetman DiamondTipstouseforbestresults Gilson Logo ID Number Pipette PIPETMAN Diamond FilterTips...

Page 9: ...older To prevent this Press and release the push button slowly and smoothly Never turn the pipette upside down Never lay the pipette on its side when there is liquidinthetip Useadedicatedstandorhang e...

Page 10: ...bly Disassemblethepipetteandreassemblethepipette correctly Ifthereisliquidinthetip holder cleanthepipette as follows Remove the tip ejector Unscrew the connecting nutandrinsethetip holder piston seala...

Page 11: ...cekit1stlevel 1tip holder 3piston sealsand3O rings F144495 F144496 F144497 F144498 F144499 Adjustmentkey F123674 F123674 F123674 F123674 F123674 Bagsof10filters F161280 Bagsof100filters F161241 If a p...

Page 12: ...12...

Page 13: ...Directives Villiers le Bel October 1st 2010 GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 VILLIERS LE BEL FRANCE Tel 33 0 1 34 29 50 00 Fax 33 0 1 34 29 50 20 PIPETMAN Fixed F2 F5 F10 F20 F25...

Page 14: ...14 NOTES...

Page 15: ...15...

Page 16: ...ed Gilson Inc World Headquarters 3000 Parmenter Street I P O Box 620027 Middleton WI 53562 0027 USA Tel 1 800 445 7661 or 1 608 836 1551 I Fax 1 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs I...

Reviews: