8
• Mold and cavity references
are marked on the collar,
ensuring the traceability. For
quality assurance purposes,
batch numbers appear on all
packages (bags and boxes).
• They form an air-tight seal
with the tip-holder, prevent-
ing the leaks that cause poor
accuracy and a lack of precision.
• PIPETMAN
diamOnd tips (except filter tips) may
be autoclaved at 121°C for 20 minutes at 0.1 mPa.
pIpEtman dIamond filter tips
Filter tips are used when sample-to-sample, sample-
to-pipette, or sample-to-operator contamination
must be avoided.
Gilson's sterilized Diamond Filter Tips are certified free
of detectable rnase, dnase, dna, rna, and protease.
pIpEtman dIamond tips to use for best results
Gilson
logo
ID Number
Pipette
PIPETMAN DIAMOND Filter Tips
Type
F2 to F25
DF30ST
Sterilized
F50 to F100
DF100ST
Sterilized
F200
DF200ST/DF300ST
Sterilized
F250 to F1000
DF1000ST
Sterilized
Pipette
PIPETMAN DIAMOND Tips
Type
F2 to F100
D200
Regular
F200
D200/D300
Regular
F250 to F1000
D1000
Regular
F5000
D5000
Regular
To ensure the best performance from your
Gilson pipette, you shoud ALWAYS use
PIPETMAN DIAMOND Tips (in accordance
with ISO8655) to complete your pipetting
system, because PIPETMAN DIAMOND Tips
were used to establish the specifications.