6

to  adjust  the  volume  setting  to 

compensate  for  a  specific  density 

or  viscosity,  slide  the  key  over  the 

push-button  until  notches  fit  into 

the grooves in the calibration sleeve 

(Figure 3), then turn the key: 
•  clockwise to decrease the volume;
•  counterclockwise to increase the 

volume;  go  above  the  desired 

value, then decrease the volume 

to reach the required setting.

Example

: When using a PIPeTmAn F100 to aspirate a par-

ticularly viscous solution, you may determine gravimetrically 

that the volume delivered is 98.8 µL, and the display reads “e 

0.2”. Therefore, you wish to increase the volume dispensed by 

1.2 µL. As the interval between each letter on the calibration 

display represents 0.4 µL for the F100 (1/10 of 4 µL), the calibra-

tion tool must be turned 3 units until the display reads “B 0.2”.

it is possible to alter the volume by more than one full 

turn in the counterclockwise direction. in this case, 

remember  to  readjust  the  same  number  of  turns 

when it is necessary to return to the original setting.

Check the new volume gravimetrically. if the volumes 

delivered are still not sufficiently close to the desired 

value, make another slight adjustment in the relevant 

direction. Be sure to change tips between volume 

setting adjustments and to pre-rinse each new tip.

When  the 

PiPetman  Fixed  is  read-

justed to compensate for a particular 

solution, the original setting (ex. 0.2) 

can  be  noted  on  the  self-adhesive 

labels supplied (Figure 4). 

Liquids  of  high  density,  vapor  pres-

sure,  or  viscosity  are  best  pipetted 

by  the  Gilson  mICrOmAn

®

  positive 

displacement pipette.

Figure 3

Figure 4

Summary of Contents for Pipetman Fixed

Page 1: ...User s Guide...

Page 2: ...sable tipsmeannoneedtohandle tips protecting the user from contamination Thetip holder theconnect ing nut and the tip ejector are autoclavable for 20 minutes at 121 C and 0 1 MPa pressure CONTENTS pag...

Page 3: ...y guaranteed using genuine PIPETMAN Diamond Tips The data given in the table conform to the ISO 8655 2 Standard Maximum Permissible Errors Model Volume Gilson ISO 8655 Reference L Systematic Random Sy...

Page 4: ...n measurements for each test volume 3 OPERATION Place a tip on the tip holder Press the tip on firmly using a slight twisting motion to ensure a positive airtight seal See Chapter 6 for the appropriat...

Page 5: ...calibrated using distilled waterandveryhighprecisionbalances Thenominal valueofthePipetmanFixedmaybeslightlyadjusted to compensate for liquids of different density or viscosity Adjustmentofthefactoryc...

Page 6: ...n displayrepresents0 4 LfortheF100 1 10of4 L thecalibra tiontoolmustbeturned3unitsuntilthedisplayreads B0 2 Itispossibletoalterthevolumebymorethanonefull turn in the counterclockwise direction In this...

Page 7: ...rate Pipetman Fixed therefore for optimum performance we recommend using PIPETMANDiamondTipswithyourPipetmanFixed PIPETMAN Diamond Tips have the Gilson logo engraved on their collar ensuring that you...

Page 8: ...oided Gilson ssterilizedDiamondFilterTipsarecertifiedfree ofdetectableRNase DNase DNA RNA andprotease pipetman DiamondTipstouseforbestresults Gilson Logo ID Number Pipette PIPETMAN Diamond FilterTips...

Page 9: ...older To prevent this Press and release the push button slowly and smoothly Never turn the pipette upside down Never lay the pipette on its side when there is liquidinthetip Useadedicatedstandorhang e...

Page 10: ...bly Disassemblethepipetteandreassemblethepipette correctly Ifthereisliquidinthetip holder cleanthepipette as follows Remove the tip ejector Unscrew the connecting nutandrinsethetip holder piston seala...

Page 11: ...cekit1stlevel 1tip holder 3piston sealsand3O rings F144495 F144496 F144497 F144498 F144499 Adjustmentkey F123674 F123674 F123674 F123674 F123674 Bagsof10filters F161280 Bagsof100filters F161241 If a p...

Page 12: ...12...

Page 13: ...Directives Villiers le Bel October 1st 2010 GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 VILLIERS LE BEL FRANCE Tel 33 0 1 34 29 50 00 Fax 33 0 1 34 29 50 20 PIPETMAN Fixed F2 F5 F10 F20 F25...

Page 14: ...14 NOTES...

Page 15: ...15...

Page 16: ...ed Gilson Inc World Headquarters 3000 Parmenter Street I P O Box 620027 Middleton WI 53562 0027 USA Tel 1 800 445 7661 or 1 608 836 1551 I Fax 1 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs I...

Reviews: