QUICK GUIDE
| MyPIPETMAN®
MYPIPETMAN® SINGLE CHANNEL
Vol.
(µL)
Maximum Permissible Errors
Gilson
ISO 8655
Systematic
Error
(µL)
Random
Error
(µL)
Systematic
Error
(%)
Random
Error
(%CV*)
Systematic
Error
(µL)
Random
Error
(µL)
Systematic
Error
(%)
Random
Error
(%CV*)
MyPIPETMAN P2 with D10 and DL10 Tips
0.2
0.5
1
2
± 0.024
± 0.025
± 0.027
± 0.030
≤
0.012
≤
0.012
≤
0.013
≤
0.014
± 12.0
± 5.0
± 2.7
± 1.5
≤
6.0
≤
2.4
≤
1.3
≤
0.7
± 0.08
± 0.08
± 0.08
± 0.08
≤
0.04
≤
0.04
≤
0.04
≤
0.04
± 40.0
± 16.0
± 8.0
± 4.0
≤
20.0
≤
8.0
≤
4.0
≤
2.0
MyPIPETMAN P10 with D10 and DL10 Tips
1
5
10
± 0.025
± 0.075
± 0.100
≤
0.012
≤
0.030
≤
0.040
± 2.5
± 1.5
± 1.0
≤
1.2
≤
0.6
≤
0.4
± 0.12
± 0.12
± 0.12
≤
0.08
≤
0.08
≤
0.08
± 12.0
± 2.4
± 1.2
≤
8.0
≤
1.6
≤
0.8
MyPIPETMAN P20 with D200 Tips
2
10
20
± 0.10
± 0.10
± 0.20
≤
0.030
≤
0.050
≤
0.060
± 5.0
± 1.0
± 1.0
≤
1.5
≤
0.5
≤
0.3
± 0.2
± 0.2
± 0.2
≤
0.1
≤
0.1
≤
0.1
± 10.0
± 2.0
± 1.0
≤
5.0
≤
1.0
≤
0.5
MyPIPETMAN P100 with D200 Tips
10
50
100
± 0.35
± 0.40
± 0.80
≤
0.10
≤
0.12
≤
0.15
± 3.5
± 0.8
± 0.8
≤
1.0
≤
0.24
≤
0.15
± 0.8
± 0.8
± 0.8
≤
0.3
≤
0.3
≤
0.3
± 8.0
± 1.6
± 0.8
≤
3.0
≤
0.6
≤
0.3
MyPIPETMAN P200 with D200 Tips
20
100
200
± 0.50
± 0.80
± 1.60
≤
0.20
≤
0.25
≤
0.30
± 2.5
± 0.8
± 0.8
≤
1.0
≤
0.25
≤
0.15
± 1.6
± 1.6
± 1.6
≤
0.6
≤
0.6
≤
0.6
± 8.0
± 1.6
± 0.8
≤
3.0
≤
0.6
≤
0.3
MyPIPETMAN P1000 with D1000 Tips
100
500
1000
± 3.0
± 4.0
± 8.0
≤
0.6
≤
1.0
≤
1.5
± 3.0
± 0.8
± 0.8
≤
0.6
≤
0.2
≤
0.15
± 8.0
± 8.0
± 8.0
≤
3.0
≤
3.0
≤
3.0
± 8.0
± 1.6
± 0.8
≤
3.0
≤
0.6
≤
0.3
MyPIPETMAN P5000 with D5000 Tips
500
2500
5000
± 12
± 15
± 30
≤
3
≤
5
≤
8
± 2.4
± 0.6
± 0.6
≤
0.6
≤
0.2
≤
0.16
± 40
± 40
± 40
≤
15.0
≤
15.0
≤
15.0
± 8.0
± 1.6
± 0.8
≤
3.0
≤
0.6
≤
0.3
MyPIPETMAN P10mL with D10mL
1000
5000
10000
± 30
± 40
± 60
≤
6
≤
10
≤
16
± 3.0
± 0.8
± 0.6
≤
0.6
≤
0.2
≤
0.16
± 60
± 60
± 60
≤
30.0
≤
30.0
≤
30.0
± 6.0
± 1.2
± 0.6
≤
3.0
≤
0.6
≤
0.3
*CV means Coefficient of Variation
Specifications
MyPIPETMAN is a high-quality pipette that has excellent
accuracy and precision, especially when used with PIPETMAN®
DIAMOND Tips. The data given in the “Gilson Maximum
Permissible Errors” table below was obtained using PIPETMAN
DIAMOND Tips. These values are guaranteed only when genuine
PIPETMAN DIAMOND Tips are used.
NOTE
The data given in this table conform to
the ISO 8655-2 standard. With a precise
pipetting technique, the P2 model can be
used to aspirate volumes as low as 0.1 μL
and the P10 model as low as 0.5 μL.