10
QUICK GUIDE
| MyPIPETMAN®
Dispensing
1.
Place the end of the tip against the inside wall of the
recipient vessel (at an angle of 10° to 40°).
2.
Press the push button slowly and smoothly to the
first
stop.
3.
Wait for at least a second, then press the push button to
the
second stop
to expel any residual liquid from the tip.
Keep the push button pressed fully, and while removing
the pipette, draw the tip along the inside surface of the
vessel.
4.
Release the push button smoothly. Eject the tip by
pressing firmly on the tip ejector button.
Second stop
First stop
Aspirate
Dispense
Prepare
Purge
Home
Rest Position
Ejecting the Tips
To avoid touching used tips, hold the pipette over the waste
container and press the tip ejector push button.
Cleaning and Decontamination
MyPIPETMAN pipettes are designed so that the parts normally
in contact with liquid contaminants can easily be cleaned and
decontaminated.
MyPIPETMAN can be sterilized by steam autoclaving at
121°C (252°F)
,
1 bar
relative pressure for
20 minutes
without
disassembly for maximum convenience as well as protection
from contamination.