7
MyPIPETMAN®
|
QUICK GUIDE
QUICK GUIDE
| MyPIPETMAN®
NOTE
To ensure good performance, always lock
the pipette before pipetting by selecting
the locking icon.
3
2
1
Figure 5
Tip ejector dual
position adapter
and extension for
P2L and P10L
Short collar
tips
Long collar
tips
180°
Fitting the Tip Ejector Adapter
Single Channel Models
A
For P2 and P10 pipettes equipped with a
stainless-steel tip ejector,
a dual-position
adapter is required to fit and eject DL10 tips
(long tips) and D10 tips (short tips).
To fit the adapter for short and long tips
(refer to Figure 5):
1.
Pull the adapter down from the
stainless steel tip ejector.
2.
Turn the adapter 180°.
3.
Refit the adapter so that the end of the
stainless steel tip ejector engages the
shorter or longer slot of the adapter.
B
For P2 and P10 pipettes equipped
with a
plastic tip ejector,
a dual-position adapter
is required to fit and eject D10 tips (short
tips).
To fit a tip ejector extension
(refer to Figure 5):
1.
Slide the extension over the tip holder.
2.
Push the extension firmly onto the
end of the tip ejector until it clicks into
place.
To remove a tip ejector extension:
1.
Gently twist the adapter.
2.
Pull it away from the pipette.
To fit
To remove
Secure Pipetting
After selecting your volume, turn the selector to the right until
the cursor reaches the locking
3
for secure pipetting.
B
A