background image

5

MyPIPETMAN®

 

|

 QUICK GUIDE

QUICK GUIDE

 

| MyPIPETMAN®

 

MYPIPETMAN® MULTICHANNEL MODELS

Maximum Permissible Errors

 Vol.
(µL)

Gilson

ISO 8655

Systematic 

error 

(µL)

Random 

error 

(µL)

Systematic 

error 

(%)

Random 

error 

(%CV*)

Systematic 

error 

(µL)

Random 

error 

(µL)

Systematic 

error 

(%)

Random 

error 

(%CV*)

MyPIPETMAN P8x10 & MyPIPETMAN P12x10 with D10 and DL10 Tips

0.5

1

5

10

± 0.08

± 0.08

± 0.20

± 0.20

 0.04

 0.05

 0.10

 0.10

± 16.0

± 8.0

± 4.0

± 2.0

 8.0

 5.0

 2.0

 1.0

± 0.24

± 0.24

± 0.24

± 0.24

 0.16

 0.16

 0.16

 0.16

± 48.0

± 24.0

± 4.8

± 2.4

 32.0

 16.0

 3.2

 1.6

MyPIPETMAN P8x20 & MyPIPETMAN P12x20 with D10L and D200 Tips

2

10

20

± 0.10

± 0.20

± 0.40

 0.08

 0.10 

 0.15

± 5.0

± 2.0

± 2.0

 4.0

 1.0

 0.75

± 0.4

± 0.4

± 0.4

 0.2

 0.2

 0.2

± 20.0

± 4.0

± 2.0

 10.0

 2.0

 1.0

MyPIPETMAN P8x200 & MyPIPETMAN P8x200 with D200 Tips

20

100

200

± 0.50

± 1.00

± 2.00

 0.25

 0.40

 0.50

± 2.5

± 1.0

± 1.0

 1.25

 0.4

 0.25

± 3.2

± 3.2

± 3.2

 1.2

 1.2

 1.2

± 16.0

± 3.2

± 1.6

 6.0

 1.2

 0.6

MyPIPETMAN P8x300 & MyPIPETMAN P12x300 with D300 Tips

20

30

150

300

± 1.00

± 1.00

± 1.50

± 3.00

 0.35

 0.35

 0.60

 1.00

± 5.0

± 3.3

± 1.0

± 1.0

 1.75

 1.17

 0.4

 0.33

± 8.0

± 8.0

± 8.0

± 8.0

 3.0

 3.0

 3.0

 3.0

± 40.0

± 26.67

± 5.33

± 2.67

 15.0

 10.0

 2.0

 1.0

*CV means Coefficient of Variation

Setting the Volume

The volume of liquid to be aspirated is set using the volume 
display. The dials are colored either black or red to indicate the 
position of the decimal point, depending on the model (refer to 
Figure 2 on page 6).

MODEL

COLOR OF VOLUMETER NUMBERS

BLACK

RED

INCREMENT

MyPIPETMAN SINGLE CHANNEL 

MyPIPETMAN P2

µL

0.01 µL

0.002 µL

MyPIPETMAN P10

µL

0.1 µL

0.02 µL

MyPIPETMAN P20

µL

0.1 µL

0.02 µL

MyPIPETMAN P100

µL

-

0.2 µL

MyPIPETMAN P200

µL

-

0.2 µL

MyPIPETMAN P1000

0.01 mL

mL

0.002 mL

MyPIPETMAN P5000

0.01 mL

mL

0.002 mL

MyPIPETMAN P10mL

mL

0.1 mL

0.02 mL

MyPIPETMAN MULTICHANNEL 

MyPIPETMAN P8x10

µL

0.1 µL

0.02 µL

MyPIPETMAN P12x10

µL

0.1 µL

0.02 µL

MyPIPETMAN P8x200

µL

-

0.2 µL

MyPIPETMAN P12x200

µL

-

0.2 µL

MyPIPETMAN P8x300

µL

-

0.2 µL

MyPIPETMAN P12x300

µL

-

0.2 µL

Summary of Contents for MyPipetman P10

Page 1: ...Quick Guide EN A U TOCLAVABL E FULLY...

Page 2: ...n the P2 and P10 models Bag of ten filters only for P5000 and P10mL models Certificate of Conformity Safety bag Multichannel Models MyPIPETMAN pipette Quick Guide Ejector spacer for D10 tips only for...

Page 3: ...AN single channel and multichannel model components Color coded push button Locking system Tip ejector button Color ring Connecting nut Tip ejector Ejector lock Ejector clip Tip holder Cover Single ch...

Page 4: ...with D200 Tips 20 100 200 0 50 0 80 1 60 0 20 0 25 0 30 2 5 0 8 0 8 1 0 0 25 0 15 1 6 1 6 1 6 0 6 0 6 0 6 8 0 1 6 0 8 3 0 0 6 0 3 MyPIPETMAN P1000 with D1000 Tips 100 500 1000 3 0 4 0 8 0 0 6 1 0 1 5...

Page 5: ...MyPIPETMAN P8x300 MyPIPETMAN P12x300 with D300 Tips 20 30 150 300 1 00 1 00 1 50 3 00 0 35 0 35 0 60 1 00 5 0 3 3 1 0 1 0 1 75 1 17 0 4 0 33 8 0 8 0 8 0 8 0 3 0 3 0 3 0 3 0 40 0 26 67 5 33 2 67 15 0...

Page 6: ...h button Light click volume selection position 1 Turn the selector until the cursor reaches the middle 2 2 Select the volume by turning the push button NOTE When selecting the volume you will hear cli...

Page 7: ...down from the stainless steel tip ejector 2 Turn the adapter 180 3 Refit the adapter so that the end of the stainless steel tip ejector engages the shorter or longer slot of the adapter B For P2 and P...

Page 8: ...Channel Models P2 P10 D10 DL10 DF10ST DFL10ST P20 D200 DF30ST P100 D200 DF100ST P200 D200 D300 DF200ST DF300ST P1000 D1000 D1200 DF1000ST DF1200ST P5000 D5000 P10mL D10mL 4 1 2 3 Multichannel Models...

Page 9: ...e tips for the best pipetting performance Ideally the pre wet includes both immersing the tip in the liquid and performing one pipetting step Pre wetting the tips helps ensure that volumes that you pi...

Page 10: ...e the push button smoothly Eject the tip by pressing firmly on the tip ejector button Second stop First stop Aspirate Dispense Prepare Purge Home Rest Position Ejecting the Tips To avoid touching used...

Page 11: ...r handling spillage or exposure to any corrosive environment abuse misuse or misapplication or from ordinary wear and tear including seal and O rings This warranty shall also be void in the event pipe...

Page 12: ...LT801608 A 2019 Gilson Inc All rights reserved gilson com contactus LT801608 B 2022 Gilson Inc All rights reserved...

Reviews: