6

set the aliquot volume, are read from top (most 
significant digit) to bottom (least significant digit). 
The marker is used to set exact or intermediate 
volumes using the scale on the bottom dial. The 
volume units (mL or µL) and position of the decimal 
point change automatically when you change the 
volume range using the selector switch.
The volumeter makes it easy to set and read any 
volume from 1 µL to 1.25 mL with accurate incre-
mental settings, as shown below.

 

DistriTip Maxi 

DistriTip Mini 

DistriTip Micro

 

Increment: 2 µL 

Increment: 0.2 µL  Increment: 0.02 µL

 

Example 1 

Example 2 

Example 3

 

0.682 mL 

68.2 µL 

6.82 µL

Setting the volume

1-  Pull up the black adjustment ring (page 5, B).
2-  With  your  Distriman  in  one  hand,  turn  the 

adjustment ring with the other hand until the 
volumeter  shows  the  required  value,  taking 
care that the marker is correctly aligned with 
the scale on the bottom dial. 

Take the following precautions in order to obtain 
maximum accuracy when changing the volumeter 
setting.
•  When 

decreasing the volume setting, turn the 

adjustment  ring  slowly  to  reach  the  required 
setting, taking care not to overshoot the mark.

•  When 

increasing  the  volume  setting,  turn 

the adjustment ring past the mark to 1/3 of a 
turn above the required setting. Then turn the 
adjustment ring slowly to decrease the volume 
setting to the required value, taking care not to 
overshoot the mark.

��

���

��

��

����

��

����

��

Summary of Contents for DISTRIMAN

Page 1: ...Other languages available on our website www gilson com...

Page 2: ...eous volatile viscous or dense into many kinds of vial tubes microplates etc Distriman ensures reliable results in any domain and in all kindsoflaboratory forawidevarietyofapplications The positive di...

Page 3: ...spare parts have been made available in case they are damaged as the result of extensive use 2 OPERATION Youshouldperformthefollowingsequenceofsteps 1 Select a DistriTip of suitable capacity 2 Fit the...

Page 4: ...aliquots varies from 118 to 121 according to the type of DistriTip used Forexample theDistriTipMicroisidealfor llinga96 well microplate Thesyringecapacityis125 L sothe Distriman can dispense at least...

Page 5: ...apacity of DistriTip tted The capacity of a DistriTip is marked on its barrel 2 Set the volume range by sliding the selector switch with your thumb to match the capacity of DistriTip tted The volume r...

Page 6: ...82 mL 68 2 L 6 82 L Setting the volume 1 Pull up the black adjustment ring page 5 B 2 With your Distriman in one hand turn the adjustment ring with the other hand until the volumeter shows the require...

Page 7: ...knobdownwards to its lowest possible position using your thumb 2 Immerse the end of the DistriTip into the liquid to be aspirated Whenaspirating ensurethatthe tip remains below the surface of the liqu...

Page 8: ...on the dampened push button ensuring that each aliquot is com pletely dispensed and that the repeatabil ity is maximized see Recommendations Ejecting the Distritip You must change the syringe before a...

Page 9: ...the side of the recipient vessel After changing the DistriTip or resetting the volume you may want to discard the rst aliquot Never try to force the volumeter beyond the recom mended volume ranges 4...

Page 10: ...ply with the requirements of the ISO 8655 Standard bytypetesting Theadjustmentiscarried outunderstrictlyde nedandmonitoredconditions ISO 8655 6 Basis of adjustment Ex Reference temperature 20 C Relati...

Page 11: ...y upwards 3 Make a note of the position and orientation of the spring K page 10 3 Pull on the retaining spring at an angle of 90 to the body of the Distriman This can be done by inserting a small scre...

Page 12: ...ag You cannot remove the red ag E page 10 until you have separated the lling knob from the piston retaining assembly In fact the act of separating these two components will free the ag Toreplacethered...

Page 13: ...2 Put all the parts into an ultrasonic bath for 15 minutes 3 Rinsethepartsthoroughlywithcleantapwater 4 Immerse all of the parts in a disinfectant bath for 15 minutes 5 Rinsethepartsthoroughlywithcle...

Page 14: ...14 NOTES...

Page 15: ...Devices GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le Bel France Tel 33 0 1 34 29 50 00 Fax 33 0 1 34 29 50 20 www gilson com Atika El Sayed Herv Ledorze General Manager Quality M...

Page 16: ...3562 0027 USA Telephone 1 800 445 7661 or 1 608 836 1551 Fax 1 608 831 4451 Gilson SAS 19 avenue des Entrepreneurs B P 145 95400 Villiers le Bel France Telephone 33 1 34 29 50 00 Fax 33 1 34 29 50 20...

Reviews: