10

The  accuracy  and  precision  figures,  given  in 
the table, are only guaranteed when you fit your 
Distriman with genuine Gilson DistriTips, of which 
the fine points results from high quality molding.

Each Distriman is engraved with its unique serial 
number, allowing you to keep track of individual 
pipettes, in accordance with GLP. The serial number 
is also printed on the warranty card.

Each pipette is inspected and validated by quali-
fied  technicians  according  to  the  Gilson  Quality 
System.

Gilson  certifies  that  its  manufactured  pipettes 
comply  with  the  requirements  of  the  ISO  8655 
Standard, by type testing. The adjustment is carried 
out under strictly defined and monitored conditions 
(ISO 8655-6):
•  Basis of adjustment, Ex.
•  Reference temperature, 20 °C
•  Relative humidity, 50 %
•  Barometric pressure, 101 kPa
•  Use of distilled water grade 3 (ISO 3696)
•  Ten measurements for each test volume, which 

are Nominal Volume, 50 % of Nominal Volume, 
and the minimum or 10 % of Nominal Volume.

5 - MAINTENANCE

Neither Distriman nor DistriTips 
may be autoclaved.

From time time it may be necessary to 
replace one of the following parts:

-  the Push-button Assembly,
-  the DistriTip Retaining Spring (K),
-  the Piston Retaining Assembly (D),
-  the Red Flag (E).

Should a persistent problem arise, 
send  the  Distriman  to  your  Gilson 
representative.

 

Summary of Contents for DISTRIMAN

Page 1: ...Other languages available on our website www gilson com...

Page 2: ...eous volatile viscous or dense into many kinds of vial tubes microplates etc Distriman ensures reliable results in any domain and in all kindsoflaboratory forawidevarietyofapplications The positive di...

Page 3: ...spare parts have been made available in case they are damaged as the result of extensive use 2 OPERATION Youshouldperformthefollowingsequenceofsteps 1 Select a DistriTip of suitable capacity 2 Fit the...

Page 4: ...aliquots varies from 118 to 121 according to the type of DistriTip used Forexample theDistriTipMicroisidealfor llinga96 well microplate Thesyringecapacityis125 L sothe Distriman can dispense at least...

Page 5: ...apacity of DistriTip tted The capacity of a DistriTip is marked on its barrel 2 Set the volume range by sliding the selector switch with your thumb to match the capacity of DistriTip tted The volume r...

Page 6: ...82 mL 68 2 L 6 82 L Setting the volume 1 Pull up the black adjustment ring page 5 B 2 With your Distriman in one hand turn the adjustment ring with the other hand until the volumeter shows the require...

Page 7: ...knobdownwards to its lowest possible position using your thumb 2 Immerse the end of the DistriTip into the liquid to be aspirated Whenaspirating ensurethatthe tip remains below the surface of the liqu...

Page 8: ...on the dampened push button ensuring that each aliquot is com pletely dispensed and that the repeatabil ity is maximized see Recommendations Ejecting the Distritip You must change the syringe before a...

Page 9: ...the side of the recipient vessel After changing the DistriTip or resetting the volume you may want to discard the rst aliquot Never try to force the volumeter beyond the recom mended volume ranges 4...

Page 10: ...ply with the requirements of the ISO 8655 Standard bytypetesting Theadjustmentiscarried outunderstrictlyde nedandmonitoredconditions ISO 8655 6 Basis of adjustment Ex Reference temperature 20 C Relati...

Page 11: ...y upwards 3 Make a note of the position and orientation of the spring K page 10 3 Pull on the retaining spring at an angle of 90 to the body of the Distriman This can be done by inserting a small scre...

Page 12: ...ag You cannot remove the red ag E page 10 until you have separated the lling knob from the piston retaining assembly In fact the act of separating these two components will free the ag Toreplacethered...

Page 13: ...2 Put all the parts into an ultrasonic bath for 15 minutes 3 Rinsethepartsthoroughlywithcleantapwater 4 Immerse all of the parts in a disinfectant bath for 15 minutes 5 Rinsethepartsthoroughlywithcle...

Page 14: ...14 NOTES...

Page 15: ...Devices GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le Bel France Tel 33 0 1 34 29 50 00 Fax 33 0 1 34 29 50 20 www gilson com Atika El Sayed Herv Ledorze General Manager Quality M...

Page 16: ...3562 0027 USA Telephone 1 800 445 7661 or 1 608 836 1551 Fax 1 608 831 4451 Gilson SAS 19 avenue des Entrepreneurs B P 145 95400 Villiers le Bel France Telephone 33 1 34 29 50 00 Fax 33 1 34 29 50 20...

Reviews: