8

2-  Press the push-button until it can 

go no farther, to dispense the first 
aliquot. This action also locks the 
adjustment ring, if it was left un-
locked after setting the volume. 

3-  Release the push-button.
4-  Repeat steps 1 to 3 for all but the 

last aliquot, which normally would 
not contain a complete aliquot. The 
red plastic flag (E, page 10), when 
completely visible, warns that it is 
time to purge and eject the syringe 
(or refill it).

5-  You  may  purge  the  syringe  at 

any time by placing the tip into a 
suitable recipient vessel and then 
pushing filling-knob downwards to 
its lowest possible position.

Operator  error  is  minimized  by  applying 
axial  pressure  on  the  dampened  push-
button,  ensuring  that  each  aliquot  is  com-
pletely  dispensed,  and  that  the  repeatabil-
ity  is  maximized  (see ‘Recommendations’).

Ejecting the Distritip

 

You  must  change  the  syringe  before  aspirating 
from a different sample or from a different liquid. 
DistriTips may discarded without having to touch 
them,  thus  avoiding  any  risk  to  the  operator. To 
remove a DistriTip, perform the following steps.
1-  If required, purge the DistriTip by placing the 

tip into a suitable recipient vessel, and then 
pushing filling-knob downwards to its lowest 
possible position.

2-  Hold  the  filling-knob  at  its  lowest  possible 

position with one hand and rotate the DistriTip 
through about 90° with the other (as if snapping-
off the syringe).

3-  When  fitting  a  new  DistriTip,  remember  to 

check  that  its  capacity  matches  the  volume 
range setting of your Distriman.

Summary of Contents for DISTRIMAN

Page 1: ...Other languages available on our website www gilson com...

Page 2: ...eous volatile viscous or dense into many kinds of vial tubes microplates etc Distriman ensures reliable results in any domain and in all kindsoflaboratory forawidevarietyofapplications The positive di...

Page 3: ...spare parts have been made available in case they are damaged as the result of extensive use 2 OPERATION Youshouldperformthefollowingsequenceofsteps 1 Select a DistriTip of suitable capacity 2 Fit the...

Page 4: ...aliquots varies from 118 to 121 according to the type of DistriTip used Forexample theDistriTipMicroisidealfor llinga96 well microplate Thesyringecapacityis125 L sothe Distriman can dispense at least...

Page 5: ...apacity of DistriTip tted The capacity of a DistriTip is marked on its barrel 2 Set the volume range by sliding the selector switch with your thumb to match the capacity of DistriTip tted The volume r...

Page 6: ...82 mL 68 2 L 6 82 L Setting the volume 1 Pull up the black adjustment ring page 5 B 2 With your Distriman in one hand turn the adjustment ring with the other hand until the volumeter shows the require...

Page 7: ...knobdownwards to its lowest possible position using your thumb 2 Immerse the end of the DistriTip into the liquid to be aspirated Whenaspirating ensurethatthe tip remains below the surface of the liqu...

Page 8: ...on the dampened push button ensuring that each aliquot is com pletely dispensed and that the repeatabil ity is maximized see Recommendations Ejecting the Distritip You must change the syringe before a...

Page 9: ...the side of the recipient vessel After changing the DistriTip or resetting the volume you may want to discard the rst aliquot Never try to force the volumeter beyond the recom mended volume ranges 4...

Page 10: ...ply with the requirements of the ISO 8655 Standard bytypetesting Theadjustmentiscarried outunderstrictlyde nedandmonitoredconditions ISO 8655 6 Basis of adjustment Ex Reference temperature 20 C Relati...

Page 11: ...y upwards 3 Make a note of the position and orientation of the spring K page 10 3 Pull on the retaining spring at an angle of 90 to the body of the Distriman This can be done by inserting a small scre...

Page 12: ...ag You cannot remove the red ag E page 10 until you have separated the lling knob from the piston retaining assembly In fact the act of separating these two components will free the ag Toreplacethered...

Page 13: ...2 Put all the parts into an ultrasonic bath for 15 minutes 3 Rinsethepartsthoroughlywithcleantapwater 4 Immerse all of the parts in a disinfectant bath for 15 minutes 5 Rinsethepartsthoroughlywithcle...

Page 14: ...14 NOTES...

Page 15: ...Devices GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le Bel France Tel 33 0 1 34 29 50 00 Fax 33 0 1 34 29 50 20 www gilson com Atika El Sayed Herv Ledorze General Manager Quality M...

Page 16: ...3562 0027 USA Telephone 1 800 445 7661 or 1 608 836 1551 Fax 1 608 831 4451 Gilson SAS 19 avenue des Entrepreneurs B P 145 95400 Villiers le Bel France Telephone 33 1 34 29 50 00 Fax 33 1 34 29 50 20...

Reviews: