13
4- Push the Coloris (as far as it will go downwards)
into position.
6 - DECONTAMINATION
You are advised to take the following steps to de-
contaminate Distriman. You can use the following
disinfectants:
a) a solution of 10% sodium hypochloride,
b) CIDEX (Johnson & Johson).
1- Disassemble the piston retaining assembly.
2- Put all the parts into an ultrasonic bath for 15
minutes.
3- Rinse the parts thoroughly with clean tap water.
4- Immerse all of the parts in a disinfectant bath
for 15 minutes.
5- Rinse the parts thoroughly with clean tap water.
6- Dry the parts using hot air (40°C to 60°C max.).
7 - SPARE PARTS
Contact your Gilson representative to order
spare parts.
Description
Reference
Push-button assembly
A, page 5
F164040
Piston retaining assembly
D, page 10
F164070
Red flag (pack of 5)
E, page 10
F164042
Retaining spring
K, page 10
F1864052