PM Menus
Submenu
Parameter
Figure 31: PM Burst Mode Menu
PM Burst Menu
PM
This menu item turns pulse modulation on and off. Pressing the adjacent softkey or
the Step Up or Step Down buttons toggles the PM state. The PM
indicator appears in
the upper right-hand corner of the display when pulse modulation is turned on.
PRI
The pulse repetition interval (PRI) of the instrument’s internal pulse modulating
signal. The PRI parameter sets the duration between similar edges of the modulating
signal, and thus, the pulse repetition interval of the individual pulses within the pulse
burst RF output. Range is 200 ns to 1 s. Each pulse in the pulse burst will have the same
PRI. Note that the PRI must be equal or greater than the pulse width plus 20 ns.
Width
The width of each pulse in a burst. Range is 10 ns to 10 ms. Each pulse in the
pulse burst will have the same pulse width.
Num of Pulses per Burst
Number of pulses in a burst. Range is 1 to 300.
Burst Period
The duration of a burst, or the repetition interval of the pulse burst. Range
is 200 ns to 10 s. Note that the burst period must be equal or greater than the PRI times
the number of pulse in the pulse burst.
Burst Mode 2/2