Ramp Operation
Giga-tronics 2500B Series Microwave Signal Generators
Page 34
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Softkey
Softkey
Softkey
4.2
Generate a Frequency-Swept Signal
This procedure describes how to set up a signal at a constant power level that sweeps linearly from a start frequency to
a stop frequency over a set duration, then repeats the sweep. The signal will be set up with the following characteristics:
Start frequency: 1.00 GHz
Stop frequency: 2.00 GHz
Power level: 1.00 dBm
NOTE:
The example procedure in this section uses specific parameters to illustrate how to setup the 2500B. You can use
this procedure for real-life situations by simply changing the parameters to fit your needs.
Table 16: Generate a Frequency Swept Signal
Generate a Frequency-Swept Signal
Step
Action
1.
Press
to display the Sweep Main menu.
2.
Press
RAMP FREQ Start/Stop Sweep menu.
3.
Select
Start Frequency, and enter 1.00 GHz using the numeric keypad (and Units
buttons),
, or
.
4.
Select
Stop Frequency, and enter 2.00 GHz using the numeric keypad (and Units
buttons),
, or
.
NOTE:
The ramp stop frequency must be set equal to or greater than the ramp start frequency.
5.
Select
Power and enter the 1.00 dBm using the numeric keypad (and Units buttons),
, or
.
6.
Select
Sweep Time, and enter 1.00 seconds using the numeric keypad (and Units
buttons),
, or
.
7.
Resolution, and select 401 using
.
Select
NOTE:
The resolution setting determines the number of discreet frequency steps that will be included
in the frequency ramp. Three resolutions are available: 401, 801, or 1601. Higher resolution settings
will result in more steps and a finer resolution ramp.
8.
If the RF ON indicator is not lit, press the RF ON button to enable the RF output. When the RF output is
enabled, the RF ON indicator is illuminated blue.
End of Procedure