Introduction
Giga-tronics 2500B Series Microwave Signal Generators
Page 6
Operation Manual, Part Number 34737, Rev C, June 30, 2011
2.5
Prepare the 2500B for Use
2.5.1
Cooling Considerations
The 2500B has an internal cooling fan. The air intake is located on the rear panel of the instrument. When placing or
installing the instrument for use, ensure there are no obstructions to the flow of air into the instrument, nor obstruction
for exhaust air flow at either side of the instrument.
2.5.2
AC Power Requirements
Table 3 below describes the power requirements and internal fuse specifications for the 2500B.
AC Line Cord:
All 2500B microwave signal generators are supplied with a 6-foot, three-wire power cord with three-
terminal polarized plug with a safety ground. If a different power cord is used, it must not exceed 3 meters (9 feet) in
order to meet safety requirements.
Table 3: 2500B AC Power and Fuse Specifications
2500B AC Power and Fuse Specifications
Parameter
Specification
AC line voltage
90 to 253 Volts.
NOTE:
The 2500B automatically adjusts to operate at
any voltage within the voltage range shown above.
No adjustments are necessary.
AC line frequency
47 to 440 Hz
Internal fuse (in rear of 2500B)
3 A, Slow-Blow, 250V, Type T
NOTE:
The procedure for replacing the fuse is in