Giga-tronics 2500B Series Microwave Signal Generators
Modulation Operation
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Page 57
Table 25: PM Menus
PM Menus
Submenu
Parameter
Figure 27: PM External Menu
PM Ext Menu
The PM External menu allows you to view and modify settings for external pulse
modulation mode, which is used when the modulating signal is provided externally. In
external PM mode, the RF output signal is pulsed by the signal that is applied to the
rear panel PULSE IN/PM TRIG IN connector. When the instrument does not include the
internal modulation generator, this is the only PM mode that is available.
Items in this submenu that can be modified are:
PM
Turns pulse modulation on and off. Press the adjacent soft key or the Step Up or
Step Down buttons toggle the PM state. The PM
indicator is displayed in the upper
right-hand corner of the display when pulse modulation is turned on.
Input Polarity
Input polarity determines the TTL level at the PULSE IN/PM TRIG IN
connector that will produce an “on” condition at the RF output. The available
selections are:
Active High
Active Low