Giga-tronics 2500B Series Microwave Signal Generators
Performance Verification
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Page 125
NOTE:
In the datasheet below, test the 2500B only at the frequencies within the range of your specific model of 2500B.
For frequencies not within your instrument’s range, enter “N/A” in the test results column(s) of the datasheet.
NOTE:
Rise Time and Fall Time are not guaranteed specifications. However, they may indicate issues with overall
performance if significantly beyond nominal specification values.
Table 80: Datasheet 10: Pulse Modulation Rise and Fall Time Test
Datasheet 10: Pulse Modulation Rise and Fall Time Test
Test
Frequency
Results
Rise
Time
Fall
Time
Level Accuracy
On/Off
Ratio (dB)
CW
Pulse
Delta
750 MHz
1
1500 MHz
1
3 GHz
6 GHz
12 GHz
18 GHz
26 GHz
39 GHz
50 GHz
Model type of 2500B series (for example, 2540B):
Serial number:
Tested by:
Date: