Password 100 (reset)
AD converter zero reset. The display shows the text
tAr
.
Make sure that the load cell simulator is zero and press the
key.
If the operation is successful, you will return to the weight display screen and you will see a value equal to
zero.
Password 200 (full scale)
Full scale of the AD converter. The display shows the indication
SPn
.
Make sure that the load cell simulator is at its full scale of 2 mV.
Use the
and
keys to modify the full scale value. Press
to set and save it.
If the operation is successful, you will return to the weight display screen and you will see a value equal to
the full scale.
Password 300 (personalized full scale)
Personalized full scale of the load cell. The display indicates the text
PrS
.
Use the
and
keys to modify the full scale value. Press
to set and save it.
Pressing
for longer than two seconds takes you back to the main screen.
Password 900 (memory reset)
Total cancellation of the settings
If the operation is successful, the display shows the indication "
rSt
" and returns to the weight display
screen.
It is necessary to switch the instrument off and on again for the changes to take effect!
Password 1234 (recovery of factory values)
In case of problems or abnormal operation of the instrument, enter this password to reload the factory set-
ting values.
This operation results in the cancellation of any present personalized settings!
If the operation is successful, the display shows the indication
FcL
and you return to the weight display
screen. It is necessary to switch the instrument off and on again for the changes to take effect.