6
PeakStore5 Manual Version 3.0
•
If necessary contact our support or contact an experienced radio or TV technician.
1.4.
Changes
According to FCC-Guidelines it is not permitted to operate the device with changes that are
not approved by GfM.
1.5.
Cables
To fulfill the thresholds of class B devices according to part 15 of the FCC-Guidelines it must
be ensured that all signal cables that are connected to the Peakanalyzer are shielded.
1.6.
Accident Prevention
It is confirmed that the delivered product is procured according to accident prevention
regulation "Elektrische Anlagen und Betriebsmittel" (BGV-A3 der Sammlung der Einzel- und
Unfallverhütungsvorschriften der gewerblichen Berufsgenossenschaften in Deutschland)
5
.
This confirmation has only the purpose to release the company of verifying the device before
the first installation (§ 5 Abs. 1, 4 der BGV-A3). Civil warranty and liability claims are not
regulated by this confirmation.
1.7.
After unpacking
After unpacking the device should be checked for mechanical damages or loose parts in the
inner. If a transport damage is occured, contact GfM and do not install the device.
1.8.
Transportation
Transport the PeakStore5 in the original package or in a package that ensures an adequate
protection. Transport damages are excluded from guarantee claims.
1.9.
Before Installation
If the PeakStore5 was carried from a cold environment to the operating area, condensation
may occur. In this situation wait until the temperature adopted to the environment and it is
absolutely dry before putting it in operation. If there is condensation due to transport or storage
wait at least 2 hours for acclimation.
1.10. Protected Earth, Shield, Power Supply
The PeakStore5 has to be connected to protected earth to fulfill the thresholds for devices of
class B according to part 15 of the FCC-Guidelines. The same counts as a requirement for the
specified technical data.
All connected signal cables to the PeakStore5 must be shielded and the shield must be
connected to protected earth. Thus the shield must have a galvanic contact with the chassis.
To connect the PeakStore5 to protected earth, an additional cable for protected earth
connection is available that can be connected to the DC IN connector for the sensor supply
(see 4.3). On the other side of the cable is an 4 mm connector that can be used to connect it
to protected earth.
5
Former VBG-4, also see http://www.bgfe.de/pages/gesetze/bgv.htm