Foot Half Rail

with Quick-Release

Hardware (Option 1)

Foot Half Rail

with Fixed Attachment

Hardware (Option 2)

OPTIONAL HALF RAIL INSTALLATION—FOOT END

(KIT ZA82900; INCLUDES TWO HALF RAILS, TWO CROSSTUBES, 

TWO MOUNTING BRACKETS, AND HARDWARE)

NOTE: Kit ZA82900, for both 80" and 76" length beds, includes attachment hardware for both 
Option 1 and Option 2 below.
IMPORTANT: Lubricate all pivot joints as needed with Aerosol White Lithium Grease.

Installation

1.  Left and right Half Rails mount to the

 

bed in a similar fashion; left or right

 

designation is based on the

 

patient’s orientation while

 

occupying the bed.

2.  To mount the half rails

 

to the bed foot end,

 

you must first attach

 

two rail brackets (#4)

 

and crosstubes (#5)

 

to the foot deck; see

 

installation instructions on

 

page 19.

3.  Align the rail pivot arm holes

 

with the mounting bracket

 

holes as shown.

4.  a.  For Quick Release (Option 1),

 

  insert the Clevis Pin ends of the

 

  Lanyard assembly (#3a) from the

 

  outside through the pivot arm and mounting

 

  bracket holes. From the inside, secure with the

 

  Hair Pin Clips on the other side of the Lanyard

 

  Assembly (#3b).

  b.  For Fixed Position (Option 2), insert the 

5

/

8

-11 x

  4.50" Cap Screws (#6) from the outside through

 

  the rail and bracket holes. From the inside,

 

  secure the screws with 

5

/

8

-11 Nylock Nuts (#7).

!

  WARNING: When assessing the Risk of Entrapment, consider

the bed, mattress, headboard, footboard, assist devices (i.e. rails 

and assist bars) and other accessories as an entire system. All 

bed systems are evaluated for full compliance to the FDA/CDRH 

“Hospital Bed System Dimensional and Assessment Guidance to 

Reduce Entrapment” guidelines.

!

  WARNING: It is extremely important to review the resident/patient’s physical and mental condition and initiate an 

appropriate individual care plan to address any entrapment risks. See page 5 for further information.

GF Health Products, Inc. - www.grahamfield.com 

Liberty Series Service Manual 999-0822-190G

LIBERTY SERIES

20

For Parts Call Medical Parts Source: 941-342-7725

Summary of Contents for Basic American Liberty Series

Page 1: ...bed model before use GF Health Products Inc www grahamfield com Liberty Series Service Manual 999 0822 190G MAY 2015 ETL UL CSA APPROVED UL 60601 1 IEC 60601 2 38 CSA 22 2 601 1 LIBERTY SERIES INSTRUC...

Page 2: ...ion available at the time of printing The most current product information can be found online at www grahamfield com Check all parts for shipping damage and test before using In case of damage DO NOT...

Page 3: ...ROL PENDANT 13 OPERATION HI LO AND DECK POSITIONS 14 OPERATION OPTIONAL PEDAL LOCK 15 SECTION B ASSIST DEVICES 16 OPTIONAL ASSIST BAR INSTALLATION HEAD END ONLY 16 OPTIONAL PIVOT ASSIST BAR INSTALLATI...

Page 4: ...y and exit Beds manufactured by Basic American Medical Products are designed for use within an institutional healthcare environment i e assisted living skilled nursing transitional care rehabilitation...

Page 5: ...pital Bed Safety Workgroup HBSW formed in 1999 GF Health Products Inc s Long Term Care Bed division Basic American Medical Products is an HBSW charter member As a result of our commitment to product s...

Page 6: ...oot pads on both the head end and foot end for any damage wear or debris Replace if needed X Check all rolling casters to ensure that they roll properly and are unobstructed X Electrical Inspection of...

Page 7: ...MENT LA MOTTIE AVENT TETE DU LIT LES COTES DE LAS TENTE OXYGENE NE DOIVENT PAS SE PROLONGER PLUS DAS QUE LA SOMMIER DU LIT ENVIRONMENTAL SPECIFICATIONS AND RFI INFORMATION OPERATING CONDITIONS Ambient...

Page 8: ...Tie CUT KEEP 5 Hand Control Pendant Pendant Holster and Coiled Power Cord 6 One Mattress Retainer 7 One Wireform Wallsaver 8 Panel Mounting Brackets and Hardware to attach Headboard and Footboard NOT...

Page 9: ...bracket tubes into the main frame rail hollow ends 6 FOR 80 BEDS Slide the mounting tubes in until the bracket tube hole aligns with the first rail hole as shown at right 7 FOR 76 BEDS Slide the mount...

Page 10: ...main frame rail hollow ends 6 FOR 80 BEDS Slide the mounting tubes in until the bracket tube hole aligns with the first rail hole as shown at right 7 FOR 76 BEDS Slide the mounting tubes in until the...

Page 11: ...as shown 3 If installing mattress retainer on the head end carefully squeeze the Mattress Retainer ends inward toward the center of the retainer and insert the ends into the proper holes see Details...

Page 12: ...necessary to attach this multi positional accessory To attach bend in the tabbed ends into the slots in the carriage rails and lock down into place There is even a position if you choose to install a...

Page 13: ...trol Pendant To order a replacement Hand Control Pendant see page 26 HAND CONTROL PENDANT OPERATION 1 HEAD DECK ANGLE UP BUTTON 2 HEAD DECK ANGLE DOWN BUTTON 3 HI LO UP BUTTON RAISE ENTIRE BED 4 HI LO...

Page 14: ...giver access during procedures and transfers The correct placement of the bed is essential for the patient s safety The bed should always be set to its lowest position when the resident patient is lef...

Page 15: ...n their down positions Liberty Bed Unlocked Position To unlock the foot end of the Liberty bed step on the green release pedal until the lock disengages The green release lever will disappear through...

Page 16: ...ng bracket holes 2 Insert the Hair Pins on the opposite end of the Lanyard Assemblies 2b into the small holes at the ends of the clevis pins Installation Fixed Attachment Head End Assist Bar Option 2...

Page 17: ...the mounting bracket on the underside of the Head Deck 1 2 From the outside insert the two 4 00 Clevis Pins 3 through the holes in the Pivoting Assist Bar bracket and Head Deck mounting bracket 3 From...

Page 18: ...nyard assembly 3a from the outside through the pivot arm and mounting bracket holes From the inside secure with Hair Pin Clips on the other side of the Lanyard Assembly 3b b For Fixed Position Option...

Page 19: ...Locate the two long U shaped Rail Mounting Brackets 4 in your kit The brackets should be oriented so that the top hole closest to the edge of the bracket is furthest away from the foot end of the bed...

Page 20: ...levis Pin ends of the Lanyard assembly 3a from the outside through the pivot arm and mounting bracket holes From the inside secure with the Hair Pin Clips on the other side of the Lanyard Assembly 3b...

Page 21: ...grain finish is facing inward toward the head deck Align the headboard holes with the Trapeze Support 2 bracket holes 4 From the back insert the four JCB Bolts 5 you removed in step 2 through the Trap...

Page 22: ...4 Slide the new footboard assembly into the main frame tubes so that the bracket second holes align with the tube first holes Use four clevis pins 6 and four hair pin clips 7 to secure the brackets t...

Page 23: ...berty bed and are attaching a second Pedal Lock you are done and can skip steps 5 8 5 Remove the two locking casters at the foot end of the carriage and replace them with two non locking casters 9 fro...

Page 24: ...ial Numbers SN12020004096 Power Frequency 120 Volt 50 60 Hz Output Rating 24V IPX4 Overall Movement Draw 2 00 Amps Classification Class 1 Type B Power Electrical Cord 18 AWG 3 Conductor Type SJT Mode...

Page 25: ...4 052SP Quantity 1 Hi Lo Motor Actuator Part 999 0848 051SP Quantity 1 Knee Foot Motor Actuator Part 999 0844 053SP Quantity 1 DO NOT use unauthorized parts accessories or adapters other than those sp...

Page 26: ...r Cable hard wired and Hardware Pack 1 Motor 999 0844 053SP OR Foot Motor Cable hard wired and Hardware Pack 1 Motor Motor Hardware Pack INCLUDES 1 Pkg 2 12mm Retaining Rings 2 Hair Pin Clips 2 Snap R...

Page 27: ...Date Price 999 0822 905G Caster Carriage Weldment 1 Unit 100 6330 062SP 3 Twin Locking Caster with 7 16 14 Stems LBT7174 1 Caster 100 6330 061SP 3 Twin Non Locking Caster with 7 16 14 Stems LBTL7174...

Page 28: ...n sheet TO ORDER A REPLACEMENT CONTROL BOX USE PART NO 999 0848 300SP Includes control box and mounting clip TO ORDER A REPLACEMENT MATTRESS RETAINER USE PART NO 999 0822 001SP TO ORDER A REPLACEMENT...

Page 29: ...ns TO ORDER A REPLACEMENT FOOT DECK USE SERVICE PACK 999 0822 930SP includes deck mounting hardware and installation instructions Bottom Pivot Channel Top Pivot Channel Foot Deck Leveler Rod Mounting...

Page 30: ...00 015 1 00 Square Tube End Cap 100 4700 015 1 00 Square Tube End Cap 100 4700 015 Main Frame Not a Service Part 1 00 Square Tube End Cap 100 4700 015 1 00 Square Tube End Cap 100 4700 015 1 00 Square...

Page 31: ...ENT YOU DAMAGE OR LOSE ANY BED CAPS OR PLUGS SEE BOTTOM OF PAGE 30 FOR ORDER NUMBERS 1 00 x 2 00 Rectangular Tube End Cap 100 4700 017 7 16 14 Nylock Nut 100 6744 001 3 00 Twin Locking Caster 7 16 14...

Page 32: ...y HAND CONTROL PENDANT DOESN T WORK 1 Check All Pendant and Cable connections and retest 2 Unplug Power Cord 3 Unplug Hand Control Pendant from the Adapter Cable Use a new Adapter Cable you know works...

Page 33: ...T GO UP AND OR DOWN 1 Check All Motor Cable connections 2 Check Motor condition cracked housing scratches bent shaft etc 3 Head and Foot Motor work but Hi Lo does not a Unplug Power Cord b Unplug Hi L...

Page 34: ...Y SERVICE This limited warranty shall only apply to defects that are reported to the Distributor from whom the Customer purchased the product within the applicable warranty period If there is not a Di...

Page 35: ...persede all prior negotiations agreements and understandings with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any representation asserti...

Page 36: ...kway Atlanta Georgia 30360 Customer Service phone 770 368 4700 fax 770 368 2386 2010 GF Health Products Inc All Rights Reserved Basic American Medical Products is a registered trademark of GF Health P...

Reviews: