SECTION F: TROUBLESHOOTING
ELECTRONIC TROUBLESHOOTING QUICK REFERENCE
The Liberty Bed uses a 3-motor system with ILCON electronics (after
September 1, 2011). To troubleshoot the Liberty Bed, follow the steps below
in the order shown.
NO POWER
1. Verify that there is power in the wall outlet.
2. Unplug Power Cord and check the following:
a. Check Power Cord for damage or pinching.
b. Check short black Control Box Cable for
damage or pinching.
c. If any cables are damaged - replace
immediately.
HAND CONTROL PENDANT DOESN’T WORK
1. Check All Pendant and Cable connections and
retest.
2. Unplug Power Cord.
3. Unplug Hand Control Pendant from the Adapter
Cable. Use a new Adapter Cable you know
works from another bed. Plug in the power cord
and test:
a. If Pendant works - replace Adapter Cable.
b. If Pendant does not work - unplug the
Pendant from Control Box. Use a new
Pendant you know works from another bed.
Plug in the power cord and test:
i. If Pendant works - replace old Pendant.
ii. If Pendant does not work - replace
Control Box.
GF Health Products, Inc. - www.grahamfield.com
Liberty Series Service Manual 999-0822-190G
LIBERTY SERIES
32
For Parts Call Medical Parts Source: 941-342-7725