![Geometrics Geode EM3D Operation Manual Download Page 22](http://html1.mh-extra.com/html/geometrics/geode-em3d/geode-em3d_operation-manual_2214700022.webp)
22
7.
Stepping cycle: This is the amount of time to go through all
the frequencies. It needs to be set to provide at least the
minimum time to run through all the frequencies. If it is not
set with enough time for all frequencies the lowest
frequencies will be truncated or dropped. If it is more than
the minimum, time will be added to the frequencies.
8.
Stepping cycle reference time; Starting reference time
9.
Local noise notch check box. When the box is checked it
will look for high amplitude noise leaking into the CSAMT
frequency acquisition band and filter out that noise. Note
this only applies to CSAMT acquisition as it cannot be done
in broad-band natural-field acquisition.
Setting the parameters in this window then clicking “Calc. Freq. Table” will generate the table of
acquisition frequencies displayed in the window. Clicking “Export Frequency Table” will create
the receiver and transmitter controller file. To export the frequency table click on the “…” box,
give it a path to the file location, and click Save.
iii.
Patch Setup: