LX-80-15 10 Hz User Manual
53
LX-80-15 10 Hz Oceanographic Radar Sensor
7. Check that the device is not operating in SDI-12 mode
The instrument can be configured to operate in SDI-12 mode. In this mode the
instrument enters low power mode unless it receives an SDI-12 command over
SDI-12 interface. While the device is in SDI-12 mode, it will not respond to
commands over RS-232 and RS-485 (Modbus) connections.
To change the instrument’s operating mode from SDI-12 mode to normal mode,
first you need to connect the Geolux Instrument Configurator to the device. To
establish a RS-232 connection with the device which operates in SDI-12 mode,
power off the device and then power it back on. After power-up, there is a one
minute period in which the device will not enter SDI-12 low power mode and it
will be possible to establish a connection over the RS-232 line. If the connection is
established within 60 seconds and as long as the Geolux Instrument Configurator
application is connected to the sensor, it will remain in operating mode. To change
between regular operating mode (
Continuous scanning
) and SDI-12 mode (
SDI-12
automatic sleep
), change the
Power management
parameter in the
Settings
tab.
The instrument
does not respond
over SDI-12
interface.
1. Make sure that the cable is connected properly to the instrument
Check that the cable M12 circular connector is firmly connected and screwed
to the instrument. It is not sufficient to simply attach the cable connector to the
instrument connector, the cable connector must be screwed into the instrument
connector.
2. Check power supply
Make sure that the power is being supplied to the instrument. A direct-current
voltage, between 9 and 27 Volts must be connected to brown (+) and white (-)
wires of the instrument cable. The power supply must be able to deliver at least
500 mA of current.
3. Check SDI-12 connection
Check and verify that the black wire from the instrument cable is connected to the
SDI-12 data line, and that the grey wire from the instrument cable is connected to
the Ground (GND) line on the SDI-12 data logger. It is important that both the data
line and the signal ground lines are connected.