LX-80-15 10 Hz User Manual
20
LX-80-15 10 Hz Oceanographic Radar Sensor
Measurement Parameters
Level Unit
The measurement unit used to report the measured level value. The default measurement unit is
millimetres. When changing the measurement unit, it’s important to make sure that other parameters
which depend on the measurement unit (such as active zone parameters) are also changed to the new
unit.
Level Offset
Changes the measurement level offset. This value should not be changed.
Active Zone Parameters
These parameters limits the operational range of the instrument. The instrument will detect water level
only within the range set by the
Active zone min.
and
Active zone max.
parameters. These parameters is
the best way to filter unwanted radar reflections from other structures and objects that are present on
the monitoring site, that could cause false instrument readings.
Active zone min.
- It is strongly recommended to set the
Active zone min.
value to the minimum possible
distance between the water and the instrument at the specific monitoring site.
Active zone max.
- It is strongly recommended to set the
Active zone max.
value to the maximum
possible distance between the water and the instrument at the specific monitoring site. Typically, this is
the distance between the instrument and the lowest point in the channel.
LX80
Active zone
ACTIVE ZONE
MAXIMUM
ACTIVE ZONE
MINIMUM